UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20230305

2023/03/05 (Sunday)

Previous Day

* Lab in-personnel: Zach

Begin another warm up and pump out
1530 Close V6 to stop circulation.
1536 V6 is now fully closed (Still pressure is rising, currently 1.5 mbar, Condenser pressure falling). Turn off 3He pump
1537 Open V3 to let mixture from Condenser side into Still side.
1540 MC(H35) = 0.62 K
1600 Close 1K Pot needle valve
1601 Turn on Still and MC heaters to 20 mW each
1602 Fill LN2 cold trap bath
1605 Close 1K Pot speedivalve, V4a, 4He pump inlet valve. Turn off 4He pump. Close recovery valve to 4He pump outlet. NOTE: remember to open recovery valve when the pump is turned back on.
1607 Unplug DCRC1
Im going to proceed more quickly than usual
1621 MC(H35)/1KPot = 3.07/2.93 K
1622 G1(fridge) = 308 mbar. Open V9 to release mixture into dump
1629 MC(H35)/1KPot = 2.88/3.09 K.
1630 Check pressure in IVC: 16 nbar, its lower at this temperature
1632 Pump out IVC with the turbo pump
1940 MC(H35) = 10.0 K. Turn off MC and Still heaters