2023/03/06 (Monday)

Previous Day

* Lab in-personnel: Zach, Himangshu

LN2 cold trap dewar filled
Begin another cooldown
1453 Begin LHe transfer @ MB = -0.9%
1457 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (open V4a, close 1K pot speedivalve and slowly open the 4He-pump inlet valve).Close the 4He-pump inlet valve once done. Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.22 bar (-23.5 inHg) on mechanical gauge.
1502 Pump out 1K Pot slowly, opening the 4He pump valve until it just starts to make noise, then opening it more when it becomes more quiet.
1504 Open 4He pump valve fully
1505 Turbo on IVC = 1 nbar, after being valved out for a min, IVC = 10 nbar, open turbo again and close valves to IVC
1508 MC(H35) = 5.47 K, cooling slowly. 1K Pot = 3.39 K / 1.3 mbar
1510 Open 1K Pot needle valve → 15° (half a unit mark). 1K Pot (P2) → 3.13 K / 6.0 mbar
1511 Open 1K Pot needle valve a little more (like 5°). 1K Pot (P2) → 2.83 K / 8.5 mbar
1518 MC temp dropping
1519 MC(H35) = 3.26 K, Open 1K Pot needle valve → 30°. 1K Pot → 2.34 K / 16 mbar.
1525 Open 1K Pot needle valve → 45°. 1K Pot → 1.95 K / 24 mbar.
1528 MC(H35)/1Kpot = 3.04/1.85 K, MC temp dropping linearly. G1/G2 = 279/262 mbar . 1K Pot(P2) = 22 mbar
1552 Ended LHe transfer @ MB = 95.9%
1609 MC(H35)/1Kpot = 2.08/1.54 K and flat. G1/G2 = 56.5/51.7 mbar . 1K Pot(P2) = 9 mbar
1610 Close V3 and V9, leaving 51.0 mbar mixture in dump. Turn on 3He pump.
1614 Start circulation through Trap 1. V6 → 12.6%. Condenser(G1) peaks at 442.2 mbar. MC(H35) → 1.48 K. Still → 1.48 K / 16 mbar.
1619 V6 → 13.0%. Condenser(G1) remains flat at 370 mbar. MC(H35) → 1.23 K. Still(P1) → 1.24 K / 12 mbar.
1622 V6 → 13.4%. Condenser(G1) remains flat at 346 mbar. MC(H35) → 1.15 K. Still(P1) → 1.13 K / 11 mbar.
1625 V6 → 13.8%. Condenser(G1) remains flat momentarily at 325 mbar. MC(H35) → 1.03 K. Still(P1) → 1.10 K / 24 mbar.
1628 V6 → 14.4%. Condenser(G1) momentarily peaks at 316.2 mbar. MC(H35) → 0.99 K. Still(P1) → 1.06 K / 26 mbar.
1634 V6 → 15.0%. Condenser(G1) peaks at 341.7 mbar but took a while to reach the peak value. MC(H35) → 0.86 K. Still(P1) → 1.11 K / 26 mbar.
1705 V6 → 100.%. Condenser(G1) peaks at 357 mbar. MC(H35) → 0.61 K. Still(P1) → 1.08 K / 26 mbar.
1731 MC(H35) = 48 mK. Plug in DCRC3.
1928 MC(H35) = 33.5 mK and pretty flat, G1 = 258 mbar and pretty flat. 1K Pot = 1.63 K / 24 mbar.
1929 Open V14 to pull the rest of the mixture out of the dump into circulation.
2150 P1 = 0.16 mbar (<1), close V14