2023/03/03 (Friday)

Previous Day

* Lab in-personnel: Zach, Ping, Himangshu

1321 MC(H35) = 5.53 K
Begin another cooldown
1331 Begin LHe transfer @ MB = 0.9%
1333 Fill LN2 cold trap dewar
1340 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (open V4a, close 1K pot speedivalve and slowly open the 4He-pump inlet valve).Close the 4He-pump inlet valve once done. Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.23 bar (-23 inHg) on mechanical gauge.
1345 Pump out 1K Pot slowly, opening the 4He pump valve until it just starts to make noise, then opening it more when it becomes more quiet.
1346 Open 4He pump valve fully
1350 Turbo on IVC = 1 nbar, after being valved out for a min, IVC = 8.6 nbar, open turbo again and close valves to IVC
1354 MC(H35) = 5.34 K, cooling slowly. 1K Pot = 3.12 K / 3.4 mbar
1355 MC temp dropping
1356 MB = 28.5%, Open 1K Pot needle valve → 15° (half a unit mark). 1K Pot (P2) → 2.73 K / 8.5 mbar
1404 MC(H35) = 3.31 K, Open 1K Pot needle valve → 30°. 1K Pot → 2.06 K / 18 mbar.
1412 Open 1K Pot needle valve → 45°. 1K Pot → 1.75 K / 30 mbar.
1415 MC(H35)/1Kpot = 2.84/1.72 K, MC temp dropping linearly. G1/G2 = 221/230 mbar . 1K Pot(P2) = 29 mbar
1437 End LHe transfer @ MB = 95.6%
1443 MC(H35)/1Kpot = 2.11/1.56 K and flat. G1/G2 = 61.5/76.4 mbar . 1K Pot(P2) = 11 mbar.
1445 Close V3 and V9, leaving 73.5 mbar mixture in dump. Turn on 3He pump.
1447 Start circulation through Trap 1. V6 → 12.6%. Condenser(G1) peaks at 435 mbar. MC(H35) → 1.54 K. Still → 1.56 K / 28 mbar.
1457 V6 → 13.0%. Condenser(G1) peaks at 312 mbar. MC(H35) → 1.22 K. Still(P1) → 1.24 K / 5.0 mbar.
1501 V6 → 13.4%. Condenser(G1) peaks at 285 mbar. MC(H35) → 1.14 K. Still(P1) → 1.17 K / 3.2 mbar.
1505 V6 → 13.8%. Condenser(G1) peaks at 282 mbar. MC(H35) → 1.04 K. Still(P1) → 1.13 K / 2.0 mbar.
1510 V6 → 14.4%. Condenser(G1) peaks at 292 mbar. MC(H35) → 0.98 K. Still(P1) → 1.10 K / 1.1 mbar.
1517 V6 → 15.0%. Condenser(G1) peaks at 306 mbar. MC(H35) → 0.94 K. Still(P1) → 1.09 K / 0.8 mbar.
1522 V6 → 100.%. Condenser(G1) peaks at 335 mbar. MC(H35) → 0.81 K. Still(P1) → 1.03 K / 0.24 mbar.
1553 MC(H35) = 41 mK and pretty flat, G1 = 248 mbar and pretty flat. 1K Pot = 1.64 K / 28 mbar.
1553 Open V14 to pull the rest of the mixture out of the dump into circulation.
1554 MC(H35) rises to 95 mK, Still maxes out then falls again, Condenser(G1) rises to 364 mbar. 1K Pot unaffected.
1611 P1 = 1.5 mbar, close V14 (this is a little early but P1 is going down slowly and im impatient)

Next Day