start › cdms › k100 › run_diary › 20230206

2023/02/06 (Monday)

Previous Day

  • Lab in-personnel: Zach, Himangshu
  • Remote: Himangshu
Status Run 77 in progress
Setup Cable 2/1/3 → diverter/nothing/nothing → DCRC50(3)/DCRC55(1)/DCRCX(4)
0940 Start MIDAS. DCRCs are setup for 2-sided 0-volt readout.
0942 Turned on DCRC1 and the DAQ programs.
0949 Turned on Pulse Viewer. Can see noise traces on PA, PB, PC and PD on DCRC1. We haven't set the SQUIDS or QET biases at this time.
0958 Turned on DCRC3 and the DAQ programs. Need to restart the towerfe and triggerfe programs multiple times to turned them on simultaneously . All channels seems to be railing.
1007 Turned on DCRC4 and the DAQ programs. Need to restart the towerfe and triggerfe programs multiple times to turned them on simultaneously . PA and PC shows noise traces but PB and PD seems to be railing. Note: PD was not showing high resistance at BOB as well
1145 Tower(RoX)/Tower(U28)/MC(H35)/Coldplate = 70.0/55.0/59.4/172.9 mK, Still(P1) = 1.29 K / 0.13 mbar / 0 mW, G1/G2 = 144.4/142.5 mbar, 1K Pot(P2) = 35° / 1.62 K / 22 mbar, MB = 68.3%
1159 Tower(RoX)/Tower(U28)/MC(H35)/Coldplate = 70.0/60.1/55.3/172.9 mK, Still(P1) = 1.28 K / 0.13 mbar / 0 mW, G1/G2 = 143.9/142.5 mbar, 1K Pot(P2) = 35° / 1.62 K / 20 mbar, MB = 67.9%
1306 Seeing large spikes in MC temp, up to 200mK earlier, seems we will have to do the same 10K pumpouts as last run
1308 Fill LN2 cold trap dewar from the 25L bomb dewar (via the 5L dewar), the bomb dewar is empty now.
1447 Switching to Tower thermometers to see if unplugging the DCRCs have any effect on the temperature. Tower(U28) = 64.2 mK
1458 A few min after unplugging the dcrcs, Tower(U28) = 54.2 mK (note that we are on the falling side of the temperature spikes
1459 temp bumped up a small amount (1mK) after plugging in dcrc3, not much change after plugging in dcrc1 and 4
1504 Seeing a rise (few mK) forming a bit after running dcrc3
1506 DCRC3 all channels seems flat and DCRC4 channel PB and PD seems flat. Doing an electrical check at the BOB to investigate this further
1715
1310 Electrical check of ZC-74 after connecting to Stripline 48 on Tower Face 3 on Cable 1
Detector: S104 Stripline 48
SC: none 50 mK
Face: 3 Date 2/06/2023
SQUET: ZC-74 Cable 1
Signal Pos. lead Gnd lead Nominal Actual Chassis
(Gnd lead)
Nominal:
< 0.1 nS
SQUID bias A/Ret 4 20 3 - 5 k 93.5 OK
SQUID bias B/Ret 19 34 93.6 OK
SQUID bias C/Ret 8 24 95.1 OK
SQUID bias D/Ret 23 39 90.9 OK
Feedback A/Ret 37 21 15 - 25 k 1.165 k OK
Feedback B/Ret 2 18 1.164 OK
Feedback C/Ret 41 25 20 NOT OK
Feedback D/Ret 22 6 1.156 k OK
QET A/Ret 3 1, 13, 14, 17, 32 60 - 80 77, 31, 21, 21, 79 OK
QET B/Ret 35 29, 33, 47, 48 33, 33, 34, 33 OK
QET C/Ret 7 5, 9, 10, 28 46, 26, 15, 22 OK
QET D/Ret 38 36, 40, 43, 44 21, 22, 21, 21 OK
QET A/QET B 3 35 1000-3000 open
QET A/QET C 3 7 open
QET A/QET D 3 38 open
QET B/QET C 35 7 open
QET B/QET D 35 38 open
QET C/QET D 7 38 5.76k
LED 1 chassis 42 1.0 - 1.1 V -
LED 2 chassis 26 -
1715 Electrical check of ZC-31 after connecting to Stripline 41 on Tower Face 5 with Cable 2
Detector: S104 Stripline 41
SC: none 50 mK
Face: 5 Date 2/06/2023
SQUET: ZC-31 Cable 2
Signal Pos. lead Gnd lead Nominal Actual Chassis
(Gnd lead)
Nominal:
< 0.1 nS
SQUID bias A/Ret 4 20 3 - 5 k 0.89M OK
SQUID bias B/Ret 19 34 98 OK
SQUID bias C/Ret 8 24 132 OK
SQUID bias D/Ret 23 39 104 OK
Feedback A/Ret 37 21 15 - 25 k 1.198 k OK
Feedback B/Ret 2 18 1.208 k OK
Feedback C/Ret 41 25 1.202 k OK
Feedback D/Ret 22 6 1.190 k OK
QET A/Ret 3 1, 13, 14, 17, 32 60 - 80 OL,40,OL, OL, OL OK
QET B/Ret 35 29, 33, 47, 48 62, 55, OL, OL OK
QET C/Ret 7 5, 9, 10, 28 57, 34, OL, OL OK
QET D/Ret 38 36, 40, 43, 44 84, 85, OL, OL OK
QET A/QET B 3 35 1000-3000 open
QET A/QET C 3 7
QET A/QET D 3 38 open
QET B/QET C 35 7 open
QET B/QET D 35 38 10.2k
QET C/QET D 7 38 open
open
LED 1 chassis 42 1.0 - 1.1 V 1.385
LED 2 chassis 26 1.452
Motor Coil Y1 169 65
Motor Coil Y2 169 54
Thermometer Pin1- Pin3 1k -
Thermometer Pin2- Pin4 1k 1167

Note: Thermometer Pin 3 seems to be open. The solution will be to make a connection between Pin 3 and Pin 4 to read it with the Lakeshore Bridge.

1402 Electrical check of ZC-26 after connecting to Stripline 43 on Tower Face 2 on Cable 3
Detector: S104 Stripline 43
SC: none 50 mK
Face: 2 Date 2/06/2023
SQUET: ZC-26 Cable 3
Signal Pos. lead Gnd lead Nominal Actual Chassis
(Gnd lead)
Nominal:
< 0.1 nS
SQUID bias A/Ret 4 20 3 - 5 k 111 OK
SQUID bias B/Ret 19 34 97 OK
SQUID bias C/Ret 8 24 99 OK
SQUID bias D/Ret 23 39 1M NOT OK
Feedback A/Ret 37 21 15 - 25 k 1.229 k OK
Feedback B/Ret 2 18 1.223 k OK
Feedback C/Ret 41 25 1.239 k OK
Feedback D/Ret 22 6 1.223 k OK
QET A/Ret 3 1, 13, 14, 17, 32 60 - 80 340, 285, 284, 280, 308 OK
QET B/Ret 35 29, 33, 47, 48 43, 43, 44, 43 OK
QET C/Ret 7 5, 9, 10, 28 47, 30, 17, 26 OK
QET D/Ret 38 36, 40, 43, 44 34, 30, 30, 30 OK
QET A/QET B 3 35 1000-3000 open
QET A/QET C 3 7 open
QET A/QET D 3 38 open
QET B/QET C 35 7 open
QET B/QET D 35 38 open
QET C/QET D 7 38 open
LED 1 chassis 42 1.0 - 1.1 V -
LED 2 chassis 26 -
1735 Switched to H35 thermometer. MC(H35) = 45.5 mK
1736 Plugged in DCRC1 and turned it on

Next Day