UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20230120

2023/01/20 (Friday)

Previous Day

  • Lab in-personnel: Zach, Himangshu, Carson
Status Run 77 preparation
Setup DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
IVC closed and Turbo pumping for a while. No tower - this was just for a leak test.
1150 Check contents of warm LN2 Trap 2: 3.4 mbar. Pump it out with the scroll pump.
1154 G1 = 1.4 mbar and stable, stop pumping on Trap 2
1156 Fill LN2 cold trap bath.
1158 Begin cryocleaning mixture through cold trap 1
1204 Open V6 to 12.0% to restart flow
1213 Open V6 to 12.3% to restart flow
Starting the procedure to open IVC. We checked the N2 gas pressure at the cylinder and lowered it to 5 PSI (it was about 10 PSI). And connected the N2 gas line to the regulator valve.
Check that the Turbo valve is closed. And Watch the gauge at the regulator value. it Reads 5 PSI. We will open it slowly till the pressure falls and stabilizes but not goes upwards.
We saw the pressure fall to 2 in the regulator gauge. We are monitoring the pressure with the gauge on the Turbo which is slowly rising from 4.6e-1 mbar towards 1 atm
1250 We let N2 leak into the ICV for 5min, the pressure gauge hooked up to the turbo maxed out at .25bar, but stopping the N2 flow and checking with the small regulator valve gauge showed that the IVC is at 2psi, a little higher than ideal
1256 We looked at the check valve on top of the fridge which is mentioned in the procedure, we are unsure of how it works though.
1305 After looking more, it seems the regulator valve gauge just needed a good tapping, it went down to -8 inHg, about matching the turbo pressure gauge.
1315 Filled the IVC with more N2 until the regulator valve gauge read 0 bar.
1355 Stop cryocleaning mixture through trap 1, pump mixture to dump.
1410 G1 = 1.2 mbar, close valves, leaving 714.5 mbar in dump.
1415 Take cold trap 1 out of the LN2 bath and put cold trap 2 in.
1420 Begin cryocleaning mixture through cold trap 2
1426 Open V6 to 12.4% to restart flow
1443 Open V6 to 12.6% to restart flow
1457 To open the IVC, we un-screwed all but two screws and inserted two screws in the two extra holes use for pushing on the vacuum seal. When the seal was broken, we heard a hissing sound of air flowing in/out? After the sound dies down, we checked the pressure reading on the regulator gauge and it went slightly left towards -1 from 0 in Hg. Is it overpressured in this case?
We remove the indium seal scrap from the bottom of the IVC and stored the IVC in the cleanroom
1617 Stop cryocleaning mixture through trap 2, pump mixture to dump.
1630 G1 = 1.3 mbar, close valves, leaving 717.9 mbar in dump
1635 Check contents of warm LN2 Trap 1: 8.6 mbar. Pump it out with the scroll pump.
1640 G1 = 2.7 mbar and stable, stop pumping on trap 1
1650 Pull cold trap 2 out of the LN2 bath and put trap 1 in
1700 Begin cryocleaning mixture through cold trap 1
1704 Open V6 to 12.4% to restart flow
1708 Open V6 to 12.6% to restart flow
1723 Open V6 to 12.7% to restart flow
1800 Stop cryocleaning mixture through trap 1, pump mixture to dump.
1810 G1 = 1.4 mbar, close valves and turn off 3He pump, leaving 713.6 mbar in dump.
1812 Pull cold trap 1 out of the LN2 bath to warm up.
xxxx We will continue the electrical check started yesterday. The pinout information: ozip
I cover the pins of Motor Coil X and piezo with kapton tape to keep them electrically insulated from accidental touch in the tower
Run 77 Tower Electrical Checkout
Device Label Unit Value
Face5pA Side1pE Ω 325.3
Face5pB Side1pB Ω 140.4
Face5pC Side1pD Ω 402.7
Face5pD Side2pE Ω 317.8
Face5pA-pB same side inter kΩ 0.627
Face5pA-pC same side inter kΩ 1.06
Face5pA-pD opp side inter kΩ 2.230
Face5pB-pC same side inter kΩ 0.711
Face5pB-pD opp side inter kΩ 1.963
Face5pC-pD opp side inter kΩ 2.182
Face5LED1 pin-to-chassis V 1.068
Face5LED2 pin-to-chassis V 0.868
Face3pA Side2pC Ω 374.5
Face3pB Side2pF Ω 512.5
Face3pC Side1pC Ω 435.6
Face3pD Side1pF Ω 403.1
Face3pA-pB same side inter kΩ 1.762
Face3pA-pC opp side inter kΩ 2.378
Face3pA-pD opp side inter kΩ 1.981
Face3pB-pC opp side inter kΩ 2.443
Face3pB-pD opp side inter kΩ 2.037
Face3pC-pD same side inter kΩ 0.811
Face2pA Side2pB Ω 140
Face2pB Side2pA Ω 86.9
Face2pC Side1pA Ω 111.6
Face2pD Side2pD Ω 385.7
Face2pA-pB same side inter kΩ 0.740
Face2pA-pC opp side inter kΩ 1.606
Face2pA-pD same side inter kΩ 1.329
Face2pB-pC opp side inter kΩ 1.734
Face2pB-pD same side inter kΩ 1.607
Face2pC-pD opp side inter kΩ 2.016
Face2pA-chassis detector shorts Ω
Face6 1-2 Motor Y 1-2 Ω 112.4
Face6 3-4 Motor Y 3-4 Ω 112.3
Face6 6-8 rox thermometer kΩ 1.039
Face6 7-9 rox thermometer kΩ 1.039
Detector: Type standard
SC: none 300 K
Face: Date 1/20/2023
SQUET: ZC-74 Who HN
Signal Pin1 Pin2 Nominal Actual Chassis
(Gnd lead)
Nominal:
< 0.1 nS
SQUID bias A/Ret 16 3 3 - 5 k 4.67 kΩ ok
SQUID bias B/Ret 27 39 4.432 k ok
SQUID bias C/Ret 19 6 4.378 k ok
SQUID bias D/Ret 30 42 4.286 k ok
Feedback A/Ret 28 41 15 - 25 k 20.98 k ok
Feedback B/Ret 14 1 19.9 k ok
Feedback C/Ret 31 44 18.42 k ok
Feedback D/Ret 17 5 18.80 k ok
QET A/Ret 2 2, 10, 11, 12, 13 14.39 k ok
QET B/Ret 4 47, 48, 49, 50 13.11 k ok
QET C/Ret 6 11, 4, 7, 8, 9 13.50 k ok
QET D/Ret 8 40, 43, 45, 46 12.88 k ok

Next Day