  • Lab in-personnel: Himangshu, Megan, Zach
Status Run 76 finished
Setup DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
IVC closed and Turbo pumping for a while. No tower - this was just for a leak test.
1100 Checked for the resources to use the new ROX thermometer. Found details about this in this link: https://www.lakeshore.com/products/categories/overview/temperature-products/cryogenic-temperature-sensors/ultra-low-temperature-rox
1230 The Motor X coils use 4 wires. The new thermometer use 2 wires. We will use the 2 wires in the cold side and switch to 4-wire measurement at the diverter board
1300 Checked for location of the thermometer. It can be fixed using the threaded tower rods Photo
1400 Fixed the Motor X arm with Kapton tape and the endstop close to the parking area Photo
1615 Drilled a small hole alongside the threaded holes on the 1D source mover carriage so that we can install it in a fixed position in the 2D source mover housing