UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20221207

2022/12/07 (Wednesday)

Previous Day

  • Lab in-personnel: Zach
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side1-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC55(1)/DCRC50(3), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe removed from lab.
NaI trigger window to 4.73-5.13 MeV
Cooled to base two days ago after thirtieth 5K pumpout.
0920 DCRC1pA baseline = -0.792 V
0928 Start -82.2 V PuBe data series ## min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
0930 HV = -91.7 V, Vtest = -21.6 mV, Ileak = -23.8 uA, Vdet = -82.2 V (90%).
DCRC1pA pulses seem strangely weak. Is it possible I got the QET bias wrong when I reset things yesterday?
0935 DCRC1pA baseline = -0.793 V
1020 DCRC1pA baseline = -0.793 V
1030 DCRC1pA baseline = -0.791 V and there are wiggles in the trace
1110 DCRC1pA baseline = -0.779 V, significantly elevated. Pulses look weaker in other channels.
about a 44 hour hold time between starting circulation and losing the phonon channels. Typical. So leaking He gas around the stripline elbow two days ago did not make things worse.
1125 Dial HV down to 0V
1128 Fill LN2 cold trap bath
1130 Close V6 to stop circulation.
1134 V6 is now fully closed (light is off on IGH, Still pressure is rising, Condenser pressure falling). Turn off 3He pump
1135 Close 1K Pot needle valve
1136 Stop DAQ programs and unplug DCRCs 1&3
1140 Open V3, releasing pressure from Condenser side to Still side
1141 Turn on Still and MC heaters to 20 mW each
1145 Close 1K Pot speedivalve, V4a, 4He pump inlet valve. Turn off 4He pump. Close recovery valve to 4He pump outlet. NOTE: remember to open recovery valve when the pump is turned back on.
1242 MC/1KPot = 3.93/4.13 K. G1(fridge) = 801 mbar.
1243 Open V9 to release pressure into the dump.
1327 MC/1KPot = 5.80/4.38 K and flat-ish
1329 Check pressure in IVC: 86 nbar
1330 Start pumping out IVC with turbo pump.
1513 Tower = 10.00 K. Turn off MC and Still heaters

Next Day