2022/12/08 (Thursday)

Previous Day

  • Lab in-personnel: Zach
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side1-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC55(1)/DCRC50(3), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe removed from lab.
NaI trigger window to 4.73-5.13 MeV
Warmed fridge yesterday for thirty-first 5K pumpout. Turbopumping IVC overnight.
1300 MB = 2.5%
1300 Turbo on IVC = 0.74 nbar. With Turbo valved out for 1 minute IVC = 7.4 nbar. Close IVC valves, for now.
We are going to do a full warmup. First: pump the mixture to the dump while the temperature is still below 10K
1315 MC/Still/1K Pot/Sorb = 5.65/6.66/>7/9.6 K
1320 Close V13a, turn on 3He pump, open V6→13.0%. This will start pumping from the fridge (both sides) and from cold LN2 Trap 1 into the dump. (V12a+speedi, V1, V3, V9 all already open)
1325 Pumping out mixture caused a little cooling, MC/Still → 5.34/6.14 K, Sorb went down then started going up, → 11.1 K and rising.
1330 Open V6 → 30% (could have done this earlier, as soon as P1 started to fall)
1338 Close Still gate valve and V1, to pump just on LN2 Trap 1 until it's empty.
1350 Isolate LN2 Trap 1 which is now pumped out to dump (close V12a and speedivalve)
1351 Continue pumping on the fridge lines (open V1 and Still Gate Valve)
1405 Stop pumping out fridge (close V1, V3, Still Gate valve, Condenser speedivalve). G1(Dump) = 714.4 mbar
Barometric pressure = 30.29“ (source is https://w1.weather.gov/data/obhistory/KMSP.html.) G2 = 714.4 mbar. Corrected dump pressure = 726.9 mbar.

Ptrue[mbar] = Pdisplay[mbar] + ( Pbar[inHg] - 29.92 ) * 33.86

1415 MC = 5.51 K. Sorb = 14.1 K
1415 Close V6. G2 = 714.7 mbar. Corrected dump pressure = 727.2 mbar.
1420 Looking more closely, G2 was still slowly rising when I closed V6, I'm going to open V6 up again → 13%
1421 Open V6 → 30%
1437 Close V6. G2 = 715.8 mbar. Corrected dump pressure = 728.3 mbar.
1445 Isolate dump (close V9) and turn off 3He pump.
1734 Recovery = 700625.59 cuft (@ 17:34:23), 16.4 cuft/hr
1745 Quick leak checker test on IVC: 1.1e-5 mbarL/s = 11000 pbarL/s, quite high background.
1750 Switch MC thermometer from H35 to Cernox (12004)
1750 MC(Cernox)/Sorb = 5.98/14.3 K