2022/12/06 (Tuesday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side1-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC55(1)/DCRC50(3), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on W side of fridge.
NaI array undergoing scan, trigger threshold varies.
Cooled to base yesterday after thirtieth 5K pumpout. Temperature spikes began around 2:00 am.
1340 MC/Coldplate = 30.2/146 mK, Still = 1.04 K / 0.080 mbar / 0 mW, G1/G2 = 190/176 mbar, 1K Pot = 45° / 1.59 K / 9.0±1.0 mbar, MB = 66.8%
1345 Fill LN2 cold trap bath
1350 Try powering DCRC1 and turning on DAQ programs. Like yesterday there's an initial MC spike to 61 mK, then it starts cooling. Channels B and C ok (have noise traces), Channel A railed like yesterday.
1400 Unplug DCRC1 from HV1 and try DCRC3 instead. Reset phonon settings appropriate for HV1. No heating but same behavior in the channels - B and C have noise traces, A railed.
1410 Try plugging DCRC2 directly into Cable 2 instead of through box. Again, B and C show traces, A railed. Has Cable2pA(Side1pE) been damaged? Note - would be better if I could zap, haven't figured that out yet.
1425 Try again to ZAP DCRC3 phonon channels, using the same ODB sequence that shows up in /MidasDAQ/online/sequencer/testzap.msl when you try to use the browser tool, which doesn't work. I think it worked this time - there were peaks in the Still pressure.
1505 Plug DCRC3 into Cable 1 to check Cable1pC(Side1pC) and Cable1pD(Side1pF). Both railed. Try zapping - again it looks like it worked, pressure spikes in still. Channels still railed.
So: maybe most of the Side1 channels were damaged (at SQUID or at QET) from the HV surge a few months ago. Only good ones are Side1pB and Side1pD (and couldn't actually get pulses in them). I guess actually the railing suggests problems at SQUID, not QET.
1335 Remove HV box. Disassemble box, remove Side1-readout board, install Side2-readout board, reassemble box. Plug HV input and test resistor leads into HV box. Reconnect cables and DCRCs as before:
Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1)
Look up phonon settings on /data/chocula/k100/midasraw/livedata/run47984.odb which from the last data series on Dec 3, enter these in ODB for DCRC1&3.
1550 Power DCRCs1&3 and turn on DAQ programs. All phonon channels look ok except DCRC3pB which can't be unrailed. (I think that's Side2pA.)
1555 Turn off DAQ programs, power down DCRCs, reset connections for DCRC3 and Cable 3, power up and check phonon channels again. Now they are all OK. good.
1615 Adjust offsets to get all phonon baselines near -0.800 V
1655 Set NaI trigger window to 4.73-5.13 MeV
1700 Remove PuBe source from lab, put in in PAN 443B storage
1719 Dial HV up to -108.1V nominal, -106.0 V actual, for 15% prebias (using RootSequencer)
1725 Dial HV down to -94V nominal, -91.7V actual. Leakage is slightly higher than I'd like but I'll leave it.
1730 HV = -91.7 V, Vtest = -22 mV, Ileak = -24.2 uA, Vdet = -82 V (89%).
1735 DCRC1pA baseline = -0.795 V
check phonon channels, had to relock DCRC3pA&D
1738 Start -82.0 V bg data series ## min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
Actually I had DCR4pA trigger on too. So there will be some random-ish triggers in this series.
1858 Start -82.1 V bg data series ## min long, NaI-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20.
Start series with the Start Button
1900 HV = -91.7 V, Vtest = -21.8 mV, Ileak = -24 uA, Vdet = -82.1 V (90%).

Next Day