UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20221128

2022/11/28 (Monday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on W side of fridge.
NaI array trigger window = 4.73 - 5.13 MeV
Cooled to base two days ago after twenty-seventh 5K pumpout.
Haven't had time to look at phonon channels this cycle, until this morning. If this cycle is typical of recent cycles we should have started to lose phonon channel sensitivity around 9:00 or 10:00 this morning.
1125 DCRC1pA baseline = -0.771 V. That's somewhat elevated, maybe. Pulses may be weaker than normal, hard to tell.
1140 MC/Coldplate = 30.7/149 mK, Still = 1.02 K / 0.060 mbar / 0 mW, G1/G2 = 108/114 mbar, 1K Pot = 45° / 1.58 K / 9.5±1.0 mbar, MB = 34.2%
1140 DCRC1pA baseline = -0.746 V and pulses definitely weak. So now it certainly looks like the He films have quenched the phonon response. Could be that the detector was relatively cool before I relocked a couple of the channels and started looking at the phonon traces, so when I first looked it didn't look as bad. Or rather, more logically, it was turning on the DAQ programs this morning that made the detector start to warm up.
1145 Fill LN2 cold trap bath
1145 Close V6 to stop circulation.
1150 V6 is now fully closed (light is off on IGH, Still pressure is rising, Condenser pressure falling). Turn off 3He pump
1150 Close 1K Pot needle valve
1155 Turn on Still and MC heaters to 20 mW each
1200 Open V3, releasing pressure from Condenser side to Still side
1200 Close 1K Pot speedivalve, V4a, 4He pump inlet valve. Turn off 4He pump. Close recovery valve to 4He pump outlet. NOTE: remember to open recovery valve when the pump is turned back on.
1300 MC/1KPot = 3.90/4.21 K. G1(fridge) = 817 mbar.
1300 Open V9 to release pressure into the dump.
1415 MC/1KPot = 6.00/4.37 K and flat
1415 Check pressure in IVC: 86 nbar
1415 Start pumping out IVC with turbo pump.
1555 Tower = 10.01 K. Turn off MC and Still heaters
At 90/60/30 minutes before reaching 10 K, MC was at 6.27/7.16/8.08 K

Next Day