start › cdms › k100 › run_diary › 20221129

2022/11/29 (Tuesday)

Previous Day

  • Lab in-personnel: Zach
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on W side of fridge.
NaI array trigger window = 4.73 - 5.13 MeV
Warmed up yesterday for twenty-eighth 5K pumpout.
1020 MC temp seems to have settled relatively high after the warmup, its now 9.68K
1022 Turbo on IVC = 0.74 nbar. With Turbo valved out for 1 minute IVC = 10 nbar
1024 Close IVC valves at top of fridge.
1047 Begin LHe transfer @ MB = 3.2%
1051 Fill LN2 cold trap dewar
1103 MB = 28.5%
1104 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (speedivalve closed). Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.42 bar on mechanical gauge.
1110 Pump out 1K Pot slowly, throttling the flow at the valve at the pump inlet, opening just until the pump gets louder, then waiting, then opening a little more, etc. With pump just a little loud, can see pressure go down at the mechanical gauge. When this is low, open valve a bit more. When P2 gets below 30 mbar open valve all the way.
1113 4He pump valve open all the way.
1114 1K Pot = 3.26 K / 2.8 mbar. G1 (fridge+dump) = 435 mbar. MC = 8.58 K.
1117 Open 1K Pot needle valve → 15°. 1K Pot → 2.88 K / 7.5 mbar
1120 1K Pot settled to 3.04 K / 5.5 mbar
1124 Open 1K Pot needle valve → 30°. 1K Pot → 2.93 K / 7.0 mbar
1125 Unplug DCRCs, Matt forgot to yesterday
1129 1K Pot settled to 2.80 K / 8.5 mbar
1130 Open 1K Pot needle valve → 45°. 1K Pot → 2.88 K / 8.5 mbar
1135 1K Pot settled to 2.76 K / 9.5 mbar
1139 Ended LHe transfer @ MB = 94.8%
1159 MC/Still/1KPot = 2.98/2.83/2.30, G1(fridge+dump) = 280 mbar, falling linearly.
1334 MC/Still/1KPot = 1.74/1.77/1.52. G1(fridge+dump) = 36 mbar, flattening.
1335 Close V3 and V9, leaving 48 mbar mixture in dump (G2)
1336 Turn on 3He pump. G1(condenser) → 45 mbar
1340 Start circulation through Trap 1. V6 → 12.6%. Condenser peaks at 176 mbar. MC(H35) → 1.48 K. Still → 1.51 K / 80 mbar
1350 V6 → 13.0%. Condenser(G1) peaks at 196 mbar. MC(H35) → 1.37 K. Still(P1) → 1.40 K / 24 mbar.
1400 V6 → 13.4%. Condenser(G1) peaks at 230 mbar. MC(H35) → 1.21 K. Still(P1) → 1.25 K / 5 mbar.
1410 V6 → 13.8%. Condenser(G1) peaks at 174 mbar (and rises more shortly after). MC(H35) → 0.85 K. Still(P1) → 1.13 K / 2.8 mbar.
1420 V6 → 14.4%. Condenser(G1) peaks at 193 mbar. MC(H35) → 0.60 K. Still(P1) → 1.12 K / 1.5 mbar.
1428 V6 → 15.0%. Condenser(G1) peaks at 192 mbar. MC(H35) → 0.37 K. Still(P1) → 1.12 K / 0.80 mbar.
1435 V6 → 100.%. Condenser(G1) peaks at 187 mbar. MC(H35) → 0.16 mK. Still(P1) → 1.07 K / 0.19 mbar.
1443 Plug in DCRCs 1 and 3
1457 Open V14 to pull the rest of the mixture out of the dump into circulation. Still peaks at 34 mbar, Condenser peaks at 189 mbar, MC peaks at 0.070 K, then all relax again. 1KPot = 1.57 K / 10±0.5 mbar.
1535 Still = 0.45 mbar. Close V14
1745 Check phonon channels. Relock some, adjust offset on DCRC1pD. All baselines are around -0.8 V. Should wait for detector temperature to stabilize before tuning these. MC temperature is rising.
2000 Tune phonon offsets to get all baselines near -0.800 V
2000 MC = 28.6 mK. Rose from 27.7 mK before turning on the DAQ programs.