UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20221127

2022/11/27 (Sunday)

Previous Day

  • Lab in-personnel: Giulia
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on W side of fridge.
NaI array trigger window = 4.73 - 5.13 MeV
Cooled to base yesterday after twenty-seventh 5K pumpout. Temperature spikes began at 1:35 am.
1545 Fill LN2 cold trap bath
1600 MC/Coldplate = 30.7/150 mK, Still = 0.98 K / 0.060 mbar / 0 mW, G1/G2 = 104/105 mbar, 1K Pot = 60° / 1.58 K / 9.5±1.0 mbar, MB = 61.7%
1605 Plugged in DCRCs 1&3

Next Day