start › cdms › k100 › run_diary › 20221126

2022/11/26 (Saturday)

Previous Day

  • Lab in-personnel: Zach
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on W side of fridge.
NaI array trigger window = 4.73 - 5.13 MeV
Warmed up yesterday for twenty-seventh 5K pumpout. Turbopumping IVC overnight.
1005 Take LN2 dewar to be filled
1029 Begin LHe transfer @ MB = 7.8% (forgot to update this before starting the transfer, so it might be a bit lower)
Also started the camera a little late, I estimated its starting value with a quick calculation of the rate.
1034 Turbo on IVC = 0.74 nbar. With Turbo valved out for 1 minute IVC = 6.3 nbar.
1036 Close IVC valves at top of fridge.
1050 MB = 27.8%
1055 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (speedivalve closed). Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.47 bar on mechanical gauge.
1058 Pump out 1K Pot slowly, throttling the flow at the valve at the pump inlet, opening just until the pump gets louder, then waiting, then opening a little more, etc. With pump just a little loud, can see pressure go down at the mechanical gauge. When this is low, open valve a bit more. When P2 gets below 30 mbar open valve all the way.
1101 4He pump valve open all the way.
1103 1K Pot = 3.18 K / 2.6 mbar. G1 (fridge+dump)=416 mbar. MC = 5.70 K.
1105 Forgot to open recovery to 4He pump before, just did
1115 Bring LN2 dewar back, fill cold trap dewar
1121 1K Pot = 3.05 K / 2.4 mbar. G1 (fridge+dump)=382 mbar. MC = 3.28 K.
1122 Open 1K Pot needle valve → 15°. 1K Pot → 2.88 K / 5.5 mbar
1127 Open 1K Pot needle valve → 30°. 1K Pot → 2.75 K / 6.5 mbar
1129 Ended LHe transfer @ MB = 95.5%
1132 Open 1K Pot needle valve → 45°. 1K Pot → 2.55 K / 9.5 mbar
1153 Set up N2 purge
1330 MC/Still/1KPot = 1.73/1.76/1.52. G1(fridge+dump) = 35 mbar, flattening.
1330 Close V3 and V9, leaving 44 mbar mixture in dump (G2)
1330 Turn on 3He pump. G1(condenser) → 43 mbar
1335 Start circulation through Trap 1. V6 → 12.6%. Condenser peaks at 196 mbar. MC(H35) → 1.39 K. Still → 1.40 K / 20 mbar
1340 V6 → 13.0%. Condenser(G1) peaks at 195 mbar. MC(H35) → 1.24 K. Still(P1) → 1.25 K / 6.5 mbar.
1345 V6 → 13.4%. Condenser(G1) peaks at 181 mbar. MC(H35) → 1.07 K. Still(P1) → 1.00 K / 1.4 mbar.
1350 V6 → 13.8%. Condenser(G1) → 155 mbar, doesn't peak. MC(H35) → 1.04 K. Still(P1) → 0.94 K / 0.6 mbar.
1355 V6 → 14.4%. Condenser(G1) peaks at 128 mbar (then again at 131 mbar after passing through the Al transition). MC(H35) → 1.02 K. Still(P1) → 0.92 K / 0.4 mbar.
1400 V6 → 15.0%. Condenser(G1) peaks at 130 mbar. MC(H35) → 0.68 K. Still(P1) → 0.88 K / 0.32 mbar.
1410 V6 → 100.%. Condenser(G1) peaks at 134 mbar. MC(H35) → 0.58 mK. Still(P1) → 0.85 K / 0.090 mbar.
1420 Open V14 to pull the rest of the mixture out of the dump into circulation. Still peaks at 32 mbar, Condenser peaks at 181 mbar, MC peaks at 0.47 K, then all relax again. 1KPot = 1.56 K / 9.5±0.5 mbar.
1440 Still = 0.85 mbar. Close V14
1440 MC = 0.12 K
1500 MC = 54 mK
1720 MC = 29.1 mK