2022/11/19 (Saturday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on W side of fridge.
NaI array trigger window = 4.73 - 5.13 MeV
Cooled to base two days ago after twenty-fourth 5K pumpout. Taking Run 68-style data overnight.
1005 MC base = 29.7 mK. Most spikes peak below 33 mK, some above 38 mK.
1005 HV = -90.4 V, Vtest = -20.3 mV, Ileak = -22.3 uA, Vdet = -81.5 V (90%).
1010 DCRC1pA baseline = -0.788 V
At 9:37 the dump sizes started to vary significantly from the norm. However the baseline doesn't seem to have risen, so it seems that helium films haven't affected the phonon channels yet.
1016 Start -81.5 V PuBe data series ## min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pA threshold 2.
Start series with the Start Button
1115 DCRC1pA baseline = -0.775 V
1145 DCRC1pA baseline = -0.776 V. Pulses seem like they may be weaker
Between 10:45 and 11:30 there was a rise and fall in the MC base, maybe also a sign that phonon channels are about to go
1215 DCRC1pA baseline = -0.773 V
1410 DCRC1pA baseline = -0.756 V and there are weird wiggles, weak pulses
1410 Close V6 to stop circulation.
1415 Dial HV down to 0V
1415 V6 is now fully closed (light is presumably off on IGH, Still pressure is rising, Condenser pressure falling). Turn off 3He pump
1415 Open V3 to let mixture from Condenser side into Still side.
1515 Close 1K Pot needle valve
1520 Fill LN2 cold trap bath
1520 Turn off DAQ programs and unplug DCRCs1&3
1535 MB = 30.8%
1540 Close 1K Pot speedivalve, V4a, 4He pump inlet valve. Turn off 4He pump. Close recovery valve to 4He pump outlet. NOTE: remember to open recovery valve when the pump is turned back on.
1545 MC/1KPot = 2.54/2.13 K. G1(fridge) = 124 mbar.
1620 MC/1KPot = 3.86/3.86 K. G1(fridge) = 749 mbar.
1620 Open V9 to release pressure into the dump.
1650 MC/1KPot = 3.46/4.37 K
1655 Check pressure in IVC: 8.6 nbar
1700 Start pumping out IVC with turbo pump.
1845 Tower = 10.00 K. Turn off MC and Still heaters
6.38 K 90 minutes
7.22 K 60 minutes
8.17 K 30 minutes
10.0 K 0 minutes