2022/11/18 (Friday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on W side of fridge.
NaI array trigger window = 4.73 - 5.13 MeV
Cooled to base yesterday after twenty-fourth 5K pumpout. MC temperature spikes began around 3:15 am
1105 Start MIDAS on DCRC2 (running CentOS7). Start DAQ programs and SQUIDtuningGUI. Everything working so far!
1105 Relock several channels. Now their baselines are all near -0.8 V
1120 Test RootSequencer at /data/chocula/cdmssoft/scdmsdaq/MidasDAQ/online/scripts/rootsequencer to change HV through DCRC2. That works too!
1300 MC temperature has increased since starting the DAQ programs and adjusting the phonon channels - 28.2 mK → 30.5 mK
1305 Adjust phonon offsets to get all baselines near -0.800 V
1306 Dial HV up to -108.1V nominal, -107.0 V actual, for 15% prebias (using RootSequencer)
1312 Dial HV down to -94V nominal, -92.6V actual
1314 Step HV down to -93V nominal, -91.6V actual
1316 Step HV down to -92V nominal, -90.5V actual
1320 HV = -90.5 V, Vtest = -20.4 mV, Ileak = -22.4 uA, Vdet = -81.6 V (90%).
1355 DCRC1pA baseline = -0.790 V
1355 MC baseline = 30.1 mK
1355 HV = -90.4 V, Vtest = -20 mV, Ileak = -22 uA, Vdet = -81.6 V (90%).
We'll try some Run68-style data (SiHV-triggered with low threshold) and take NaI data with this new threshold window.
1404 Start -81.6 V PuBe data series ## min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pA threshold 2.
Start series with the Start Button
1410 Data seems to be writing correctly. MIDAS seems to be functioning.
1435 MC/Coldplate = 30.0/153 mK, Still = 1.12 K / 0.100 mbar / 0 mW, G1/G2 = 148/137 mbar, 1K Pot = 60° / 1.57 K / 9.0±1.0 mbar, MB = 65.2%
1440 Fill LN2 cold trap bath
1850 DCRC1pA baseline = -0.784 V
1850 Start -81.6 V PuBe data series ## min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pA threshold 2.
Start series with the Start Button
1850 HV = -90.5 V, Vtest = -20.4 mV, Ileak = -22.4 uA, Vdet = -81.6 V (90%).