2022/11/20 (Sunday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on W side of fridge.
NaI array trigger window = 4.73 - 5.13 MeV
Warmed up yesterday for twenty-fifth 5K pumpout. Turbopumping IVC overnight.
1041 Begin LHe transfer @ MB = 5.0%
1045 Turbo on IVC = 0.74 nbar. With Turbo valved out for 1 minute IVC = 7.4 nbar.
1045 MC = 5.76 K. For some reason the MC cooled off a lot more this time than the last few times.
1050 Fill LN2 cold trap bath
1055 MB = 29.4%
1055 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (speedivalve closed). Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.33 bar on mechanical gauge.
1100 Pump out 1K Pot slowly, throttling the flow at the valve at the pump inlet, opening just until the pump gets louder, then waiting, then opening a little more, etc. With pump just a little loud, can see pressure go down at the mechanical gauge. When this is low, open valve a bit more. When P2 gets below 30 mbar open valve all the way.
1105 4He pump valve open all the way.
1105 1K Pot = 3.15 K / 3.2 mbar. G1 (fridge+dump)=411 mbar. MC = 5.68 K.
1125 1K Pot = 2.93 K / 3.2 mbar. G1 (fridge+dump)=352 mbar. MC = 3.22 K.
1135 1K Pot = 2.87 K / 3.2 mbar. G1 (fridge+dump)=339 mbar. MC = 3.19 K.
1135 Open 1K Pot needle valve → 15°. 1K Pot → 2.74 K / 5.0 mbar
1140 Open 1K Pot needle valve → 30°. 1K Pot → 2.59 K / 6.5 mbar
1143 End LHe transfer @ MB = 93.6%
1145 Open 1K Pot needle valve → 45°. 1K Pot → 2.37 K / 9.5 mbar
1240 MC/Still/1KPot = 2.16/1.94/1.52. G1(fridge+dump) = 60 mbar and getting flat.
1245 Close V3 and V9, leaving 57 mbar mixture in dump (G2)
1245 Turn on 3He pump. G1(condenser) → 76 mbar
1245 Start circulation through Trap 1. V6 → 12.6%. Condenser peaks at 401 mbar. MC(H35) → 1.66 K. Still → 1.60 K / 50 mbar
1250 V6 → 13.0%. Condenser(G1) peaks at 365 mbar. MC(H35) → 1.42 K. Still(P1) → 1.43 K / 9.5 mbar.
1252 V6 → 13.4%. Condenser(G1) → 315 mbar, doesn't peak. MC(H35) → 1.19 K. Still(P1) → 1.21 K / 3.6 mbar.
1255 V6 → 13.8%. Condenser(G1) → 267 mbar, doesn't peak. MC(H35) → 1.03 K. Still(P1) → 1.09 K / 1.8 mbar.
1257 V6 → 14.4%. Condenser(G1) → 232 mbar, doesn't peak. MC(H35) → 0.93 K. Still(P1) → 0.97 K / 0.6 mbar.
1300 V6 → 15.0%. Condenser(G1) → 192 mbar, doesn't peak. MC(H35) → 0.88 K. Still(P1) → 0.94 K / 0.40 mbar.
1300 1K Pot = 1.53 K / 8.0 mbar
1302 Open 1K Pot needle valve → 60°. 1K Pot → 1.53 K / 8.0 mbar, no change really
1305 V6 → 100.%. Condenser(G1) peaks at 152 mbar. MC(H35) → 0.74 mK. Still(P1) → 0.89 K / 0.13 mbar.
1310 Open V14 to pull the rest of the mixture out of the dump into circulation. First open for several seconds then close again - Still peaks at 150 mbar, Condenser peaks at 160 mbar. Open and close again: Still peaks at 150 mbar, Condenser peaks at 168 mbar. Again: Still peaks at 160 mbar, Condenser peaks at 174 mbar. Open V14 and leave it open: Still peaks at 190 mbar, Condenser peaks at 231 mbar.
1315 MC = 0.32 K. 1K Pot = 1.57 K / 12 mbar.
1315 Plug in DCRCs 1 and 3
1335 Still = 0.90 mbar. Close V14
1355 MC = 40.0 mK
1505 MC = 28.8 mK
2015 Can't start a MIDAS run with DCRC1 enabled, DAQ programs crash. Will have to try powercycling it.