start › cdms › k100 › run_diary › 20221115

2022/11/15 (Tuesday)

Previous Day

  • Lab in-personnel: Giulia
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on W side of fridge.
NaI array trigger window = 4.73 - 5.13 MeV
Cooled down yesterday after twenty-third 5K pumpout. Temperature spikes began at 4:40 am.
0845 HV = -90.9 V, Vtest = -20.8 mV, Ileak = -22.9 uA, Vdet = -81.7 V (90%).
0845 Start -81.7 V PuBe data series 33 min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
0910 MC/Coldplate = 29.7/153 mK, Still = 1.13 K / 0.095 mbar / 0 mW, G1/G2 = 151/135 mbar, 1K Pot = 60° / 1.58 K / 10.5±0.5 mbar, MB = 69.8%
0920 Fill LN2 cold trap bath
0919 Start -81.7 V PuBe data series ## good min long, NaI-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20.
Start series with the Start Button
1520 DCRC1pA baseline = -0.791 V
2105 DCRC1pA baseline = -0.789 V
2105 HV = -90.8 V, Vtest = -20.8 mV, Ileak = -22.9 uA, Vdet = -81.7 V (90%).
2109 Start -81.7 V PuBe data series ## min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
2201 Start -81.7 V PuBe data series ## good min long, NaI-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20.
Start series with the Start Button

Next Day