2022/11/14 (Monday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on W side of fridge.
NaI array trigger window = 4.73 - 5.93 MeV
Warmed up yesterday for twenty-third 5K pumpout. Turbo pumping IVC overnight.
0932 Begin LHe transfer @ MB = 3.6%
0940 Turbo on IVC = 0.86 nbar. With Turbo valved out for 1 minute IVC = 10 nbar.
0940 Close IVC valves at top of fridge.
0945 Fill LN2 cold trap bath
0950 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (speedivalve closed). Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.38 bar on mechanical gauge.
0950 Pump out 1K Pot slowly, throttling the flow at the valve at the pump inlet, opening just until the pump gets louder, then waiting, then opening a little more, etc. With pump just a little loud, can see pressure go down at the mechanical gauge. When this is low, open valve a bit more. When P2 gets below 30 mbar open valve all the way.
0955 4He pump valve open all the way.
0955 1K Pot = 3.18 K / 3.4 mbar. G1 (fridge+dump)=426 mbar. MC = 9.59 K.
MC didn't cool very much overnight after turning off the heaters this time. Hopefully this won't be a problem…
1010 1K Pot = 3.10 K / 3.4 mbar. G1 (fridge+dump)=396 mbar. MC = 9.15 K.
1030 1K Pot = 2.99 K / 3.4 mbar. G1 (fridge+dump)=376 mbar. MC = 5.87 K.
1030 Open 1K Pot needle valve → 15°. 1K Pot → 2.89 K / 5.5 mbar
1040 MC = 3.26 K. G1(fridge+dump) = 373 mbar.
1041 End LHe transfer @ MB = 95.6%
1055 1K Pot = 2.54 K / 5.7±0.3 mbar
1055 Open 1K Pot needle valve → 30°. 1K Pot → 2.37 K / 6.7±0.3 mbar
1110 Open 1K Pot needle valve → 45°. 1K Pot → 1.87 K / 9.5±0.5 mbar
1120 MC = 2.66 K. G1(fridge+dump) = 156 mbar.
1125 Open 1K Pot needle valve → 60°. 1K Pot → 1.71 K / 12 mbar
1720 MC/Still/1KPot = 1.68/1.73/1.59. G1(fridge+dump) = 32 mbar and flat.
1725 Close V3 and V9, leaving 23 mbar mixture in dump (G2)
1725 Turn on 3He pump. G1(condenser) → 38 mbar
1725 Start circulation through Trap 1. V6 → 12.6%. Condenser peaks at 250 mbar. MC(H35) → 1.40 K. Still → 1.45 K / 32 mbar
1735 V6 → 13.0%. Condenser(G1) peaks at 269 mbar. MC(H35) → 1.17 K. Still(P1) → 1.22 K / 4.5 mbar.
1740 V6 → 13.4%. Condenser(G1) peaks at 220 mbar. MC(H35) → 1.0 K. Still(P1) → 1.18 K / 3.0 mbar.
1800 V6 → 13.8%. Condenser(G1) peaks at 190 mbar. MC(H35) → 0.47 K. Still(P1) → 1.16 K / 1.8 mbar.
1810 V6 → 14.4%. Condenser(G1) peaks at 192 mbar. MC(H35) → 0.27 K. Still(P1) → 1.17 K / 0.8 mbar.
1820 V6 → 15.0%. Condenser(G1) peaks at 179 mbar. MC(H35) → 0.21 K. Still(P1) → 1.15 K / 0.55 mbar.
1825 V6 → 100.%. Condenser(G1) peaks at 182 mbar. MC(H35) → 93 mK. Still(P1) → 1.10 K / 0.16 mbar.
1830 Open V14 to pull the rest of the mixture out of the dump into circulation. Still peaks at 24 mbar, Condenser peaks at 204 mbar, MC peaks at 99 mK, then all relax again. 1K → 1.58 K / 12±1 mbar.
1900 Still = 0.55 mbar. Close V14
1905 MC = 30.4 mK
1930 MC = 27.6 mK
2010 MC = 27.3 mK
2025 Plug in DCRCs 1&3
2030 Check phonon channels. Relock and adjust some of the offsets to get baselines ~ -0.8 V
2120 Adjust phonon offsets to get all baselines near -0.800 V
2120 MC = 28.9 mK
2127 Dial HV up to -108.1V nominal, -106.2 V actual, for 15% prebias (using RootSequencer)
2130 Set NaI array trigger window to 4.73 - 5.13 MeV
2133 Dial HV down to -94V nominal, -91.9V actual
2135 Step HV down to -93V nominal, -90.8V actual
2135 HV = -90.8 V, Vtest = -20.6 mV, Ileak = -22.7 uA, Vdet = -81.8 V (90%).
2140 DCRC1pA baseline = -0.789 V
2141 Start -81.8 V PuBe data series ## min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
2220 DCRC1pA baseline = -0.797 V
2223 Start -81.8 V PuBe data series ## good min long, NaI-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20.
Start series with the Start Button

Next Day