start › cdms › k100 › run_diary › 20221116

2022/11/16 (Wednesday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on W side of fridge.
NaI array trigger window = 4.73 - 5.13 MeV
Cooled down two days ago after twenty-third 5K pumpout. Taking -82V data overnight with NaI window around 5 MeV
1000 HV = -90.8 V, Vtest = -21 mV, Ileak = -23.1 uA, Vdet = -81.6 V (90%).
1000 DCRC1pA baseline = -0.790 V
1000 MC base = 29.6 mK with most spikes peaking below 33 mK, some above 34 mK
1003 Start -81.6 V PuBe data series ## min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
1048 Start -81.7 V PuBe data series ## good min long, NaI-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20.
Start series with the Start Button
1135 DCRC1pA baseline = -0.783 V
1135 MC/Coldplate = 30.0/151 mK, Still = 1.13 K / 0.100 mbar / 0 mW, G1/G2 = 141/128 mbar, 1K Pot = 60° / 1.58 K / 11.5±0.5 mbar, MB = 33.2%
1140 Fill LN2 cold trap bath
1335 DCRC1pA baseline = -0.742 V. Clearly elevated - He films are now messing up the phonon channels. Time to warm up.
1345 Dial HV down to 0V
1350 Turn off DAQ programs. Also stop MIDAS on cdms1 so we can test it on cdms2
1355 Close V6 to stop circulation.
1400 V6 is now fully closed (light is off on IGH, Still pressure is rising, Condenser pressure falling). Turn off 3He pump
1400 Open V3 to let mixture from Condenser side into Still side.
1400 Close 1K Pot needle valve
1405 MC = 0.38 K
1405 Turn on Still and MC heaters to 20 mW each
1435 Close 1K Pot speedivalve, V4a, 4He pump inlet valve. Turn off 4He pump. Close recovery valve to 4He pump outlet. NOTE: remember to open recovery valve when the pump is turned back on.
1510 MC/1KPot = 3.86/3.99 K. G1(fridge) = 748 mbar.
1510 Open V9 to release pressure into the dump.
1525 MC/1KPot = 6.17/6.37 K, flat
1525 Check pressure in IVC: 0.086 ubar, strangely low. Start pumping out IVC with turbo pump.
1525 MB = 25.9%
1900 Tower = 10.01 K. Turn off MC and Still heaters

Next Day