2022/11/13 (Sunday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on W side of fridge.
NaI array trigger window = 4.73 - 5.93 MeV
Cooled to base two days ago after twenty-second 5K pumpout.
Taking -82V data overnight. Trying lower NaI trigger window, targeting a different strong line in the Si (n,gamma) spectrum
1030 MC base = 30.0 mK. Most spikes peak below 33 mK, some above 35 mK
1035 HV = -91.8 V, Vtest = -22.3 mV, Ileak = -24.5 uA, Vdet = -82 V (89%).
1040 DCRC1pA baseline = -0.771 V. Seems like it may already be elevated…
like last time it looks like the phonon channels are affected by helium films around 44-45 hours after cooling to base (or maybe 43). Shorter than it was in the past.
1159 Start -82.0 V PuBe data series ## min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
1200 DCRC1pA baseline = -0.759 V. Seems even more elevated…
Other channels also look like they have elevated baselines and there may be wiggles in the traces from helium films.
1235 DCRC1pA baseline = -0.728 V. Seems even more elevated… Pulses look weak too.
1250 Dial HV down to 0V after stopping data series. Turn off DAQ programs.
1255 Close V6 to stop circulation.
1305 V6 is now fully closed (light is presumably off on IGH, Still pressure is rising, Condenser pressure falling). Turn off 3He pump
1205 Open V3 to let mixture from Condenser side into Still side.
1450 Close 1K Pot needle valve
1450 MC = 0.57 K
1450 Turn on Still and MC heaters to 20 mW each
1455 Fill LN2 cold trap bath
1455 MB = 28.9 %
1530 Close 1K Pot speedivalve, V4a, 4He pump inlet valve. Turn off 4He pump. Close recovery valve to 4He pump outlet. NOTE: remember to open recovery valve when the pump is turned back on.
1605 MC/1KPot = 3.82/4.08 K. G1(fridge) = 717 mbar.
1605 Open V9 to release pressure into the dump.
1610 MC/1KPot = 3.24/4.04 K. Gonna start pumping out a bit earlier than usual.
1615 Check pressure in IVC: 0.01 ubar. Start pumping out IVC with turbo pump.
1705 MC/1KPot/Sorb = 3.40/4.39/5.5 K.
1845 Tower = 10.66 K. Turn off MC and Still heaters