2022/09/29 (Thursday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe source in storage in PAN443.
NaI trigger window: 7.6-11.2 MeV
Cooled to base two days ago after tenth 5K pumpout.
1500 MC base 29.3 mK with most spikes peaking below 33 mK, some above 40 mK
1505 MC/Coldplate = 29.3/150 mK, Still = 1.12 K / 0.110 mbar / 0 mW, G1/G2 = 116/104 mbar, 1K Pot = 45° / 1.58 K / 12±1 mbar, MB = 65.0%
1510 Fill LN2 cold trap bath
1510 Phonon baselines are all around -0.796 V, very slightly up from yesterday
1520 Checking to see how low a threshold we can set. First looking at data it looks like Side2pC = DCRC1pA is the channel with lowest noise. Setting that threshold down to 3 gives all good pulses. Setting it to 2 there's a lot of noise triggers, but there are still plenty of pulses showing up in the right place. So that's probably roughly similar to the low threshold setting in Run68.
1526 Start 0 V bg data series ## min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pA threshold 2.
Start series with the Start Button
1649 Dial HV up to -109.25V nominal, -107.2V actual, for 15% prebias (using RootSequencer)
1626 Dial HV down to -95V nominal, -93.0V actual
1627 HV = -93 V, Vtest = -21.7 mV, Ileak = -23.9 uA, Vdet = -83.4 V (90%).
1628 Start -83.4 V bg data series ## min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pA threshold 2.
Start series with the Start Button
abort series because DCRC2 was enabled, let's disable it to keep data cleaner
1629 Start -83.4 V bg data series ## min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pA threshold 2.
Start series with the Start Button