start › cdms › k100 › run_diary › 20220927

2022/09/27 (Tuesday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe source in storage in PAN443.
NaI trigger window: 7.6-11.2 MeV
Tenth 5K pumpout last night.
1245 MC/1KPot/Sorb = 5.32/6.52/8.3 K. G1(fridge) = 543 mbar
1245 Turbo on IVC = 1.4 nbar. With Turbo valved out for 1 minute IVC = 12 nbar.
1245 Close IVC valves on top of fridge
1250 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (speedivalve closed). Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.24 bar on mechanical gauge.
1250 Pump out 1K Pot in stages, alternating 1K Pot Speedivalve and V4a, opening V4a in brief spurts to keep 4He pump pressure low (below 5 mbar).
1255 Start pumping 1K Pot continuously
1255 MC/1KPot/Sorb = 5.32/3.44/4.3 K
1315 MC drops quickly to 3.4 K
1340 MC/1KPot/Sorb = 3.35/3.08/4.0 K. G1(fridge) = 421 mbar and falling
1400 MC/1KPot/Sorb = 3.23/2.88/4.0 K. G1(fridge) = 367 mbar. MC, 1KPot, G1 falling linearly.
1450 MC/1KPot = 2.98/2.54. G1(fridge) = 262 mbar. All still falling.
1505 1K Pot = 0° / 2.48 K / 2.2 mbar
1505 Open 1K Pot needle valve → 15°. 1K Pot → 2.28 K / 5.5 mbar
1505 MC = 2.89 K. G1(fridge) = 225 mbar
1515 Open 1K Pot needle valve → 30°. 1K Pot → 1.95 K / 8.0 mbar
1515 MC = 2.76 K. G1(fridge) = 187 mbar
1520 Fill LN2 cold trap bath
1520 Open 1K Pot needle valve → 45°. 1K Pot → 1.69 K / 12 mbar
1602 Sudden drop in MC from 2.07→1.99 K, looks like it got past the lambda point
1605 MC = 1.95 K. G1(fridge) = 53 mbar, pretty flat
1610 1K Pot = 1.54 K / 7.0 mbar
1610 Close V3 and V9, leaving 40 mbar mixture in dump (G2)
1610 Turn on 3He pump. G1(condenser) → 63 mbar.
1610 Start circulation through Trap 1. V6 → 12.6%. Condenser peaks at 303 mbar. MC(H35) → 1.51 K. Still → 1.55 K / 60 mbar.
1615 V6 → 13.0%. Condenser(G1) peaks at 250 mbar. MC(H35) → 1.21 K. Still(P1) → 1.25 K / 6.0 mbar.
1620 1K Pot = 1.59 K / 12.0 mbar
1620 V6 → 13.4%. Condenser(G1) peaks at 196 mbar. MC(H35) → 1.09 K. Still(P1) → 1.16 K / 3.8 mbar.
1625 V6 → 13.8%. Condenser(G1) peaks at 177 mbar. MC(H35) → 0.88 K. Still(P1) → 1.13 K / 2.4 mbar.
MC kinked around 1.2 and 1.1 K, Al Tc.
1635 V6 → 14.4%. Condenser(G1) peaks at 176 mbar. MC(H35) → 0.36 K. Still(P1) → 1.15 K / 0.95 mbar.
1645 V6 → 15.0%. Condenser(G1) peaks at 166 mbar. MC(H35) → 0.24 K. Still(P1) → 1.14 K / 0.60 mbar.
1650 V6 → 100%. Condenser(G1) peaks at 164 mbar. MC(H35) → 0.14 K. Still(P1) → 1.08 K / 0.17 mbar.
1720 MC = 32.5 mK. Condenser(G1) = 118 mbar. Still = 0.110 mbar.
1720 Open V14 to pull the rest of the mixture out of the dump into circulation. Still peaks at 140 mbar, Condenser peaks at 193 mbar, MC peaks at 91 mK, then all relax again. 1K Pot unaffected.
1745 Still = 0.95 mbar.
Close V14
1745 Plug in DCRCs 1&3