2022/09/26 (Monday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe source in storage in PAN443.
NaI trigger window: 7.6-11.2 MeV
Cooled to base three days ago after ninth 5K pumpout. Took -82.5V data overnight.
0905 Check phonon channels: no pulses, elevated baselines, most above -0.7 V
0910 Dial HV down to 0V
0920 Phonon baselines for DCRC1pA,B,D,DCRC3pA,B,D = -0.725, -0.682, -0.711, -0.648, -0.618, -0.679 V = Side2pC,F,E,B,A,D
MC baseline temperature overnight: began around 29.5 mK, @ 19:30: 29.9 mK, @ 24:15: 30.2 mK, @ 1:55: 30.6 mK, @ 8:10: 31.0 mK
0925 Close V6 to stop circulation
0935 Turn off 3He pump
0940 Open V3; G1 falls, so this let mixture from Condenser side over to Still side.
1230 Close 1K Pot needle valve
1250 Close 1K Pot speedivalve, V4a, 4He pump inlet valve. Turn off 4He pump. Close recovery valve to 4He pump outlet. NOTE: remember to open recovery valve when the pump is turned back on.
1250 Turn off DAQ programs and unplug DCRCs 1&3
1255 Turn on Still and MC heaters to 20 mW each
1410 G1(fridge) = 855 mbar.
Open V9 to release mixture into dump.
1500 MC/1Kpot/Sorb = 5.77/4.37/5.5 K and pretty flat
1505 Check pressure in IVC: 0.10 ubar.
1515 Wedge a couple of washers between the knurled knobs of the stripline elbow on the side near the center of the fridge.
1515 Star pumping out IVC with turbo pump
1527 Begin LHe transfer @ MB = 33.4%
1622 End LHe transfer @ MB = 95.5%
1650 Fill LN2 cold trap bath
1705 MC/1Kpot/Sorb = 10.1/5.25/6.6 K
Turn off MC and Still heaters