UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20220918

2022/09/18 (Sunday)

Previous Day

  • Lab in-personnel: Matt, Giulia
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on West side of lead/poly shield.
NaI trigger window: 7.6-11.2 MeV
Cooled to base a couple days ago after seventh 5K pumpout.
Taking -80V data overnight. DAQ crash at 2:30 am, no data after that.
1145 As of this morning I can't log into cdms1. So we can't access/restart the DAQ.
1435 Figured out how to reduce the bias to 0 without the DAQ: telnet commands to DCRC2. See setting_up_dcrc_boards. Also see RevC.0 Register Map. Connected from cdms3 with telnet 128.101.21.162 5001. Did rd 5A (qI charge bias) - this was 3289. Doing wr 5A 3290 caused the HV to drop. So I stepped this up manually (in hexadecimal); at 3A55 I got the HV down to -12 mV
1545 MC/Coldplate = 29.4/150 mK, Still = 1.12 K / 0.075 mbar / 0 mW, G1/G2 = 130/128 mbar, 1K Pot = 45° / 1.59 K / 13±1 mbar
By now or soon helium films will kill the phonon channels, so let's go ahead and start warming up.
1550 Close V6 to stop circulation.
1555 Turn off 3He pump
1615 Close 1K Pot needle valve
1630 Turn on Still and MC heaters to 20 mW each
1644 Begin LHe transfer @ MB = 29.3%
1655 Unplug DCRCs 1&3
1700 Fill LN2 cold trap bath
1700 Close 1K Pot speedivalve, V4a, 4He pump inlet valve. Turn off 4He pump. Close recovery valve to 4He pump outlet. NOTE: remember to open recovery valve when the pump is turned back on.
1700 Open V3; G1 rises, so apparently Still side was at higher pressure than Condenser side.
1731 End LHe transfer @ MB = 95.7%
1750 MC/1Kpot = 3.87/4.23 K and rising. Fridge (G1) = 755 mbar and rising
1750 Open V9 to release mixture into dump. MC temperature drops, then starts rising again
1815 MC reaches 3.5 K then starts rising much faster
1940 MC/1Kpot/Sorb = 6.26/4.33/5.4 K and flat
1940 Check pressure in IVC: 0.086 ubar. Star pumping out IVC with turbo pump
MC temperature starts rising again
2120 MC = 10.0 K.
Turn off MC and Still heaters

Next Day