2022/09/17 (Saturday)
- Lab in-personnel: Giulia
| Status | Run 76 in progress |
|---|---|
| Setup | Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4) Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors. PuBe behind lead wall on West side of lead/poly shield. |
| NaI trigger window: 7.6-11.2 MeV | |
| Cooled to base yesterday after seventh 5K pumpout. Temperature spikes started at 5:50 am | |
| Taking -80V data overnight | |
| 1039 | Start -80.2 V PuBe data series 103 min long, SiHV-triggered, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10. Start series with the Start Button |
| 1045 | HV = -88 V, Vtest = -17.7 mV, Ileak = -19.4 uA, Vdet = -80.2 V (91%). |
| 1225 | Start -80.3 V PuBe data series 531 min long, NaI-triggered, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20. Start series with the Start Button |
| 1225 | HV = -88 V, Vtest = -17.4 mV, Ileak = -19.1 uA, Vdet = -80.3 V (91%). |
| 1700 | MC/Coldplate = 31.3/149 mK, Still = 1.12 K / 0.075 mbar / 0 mW, G1/G2 = 126/129 mbar, 1K Pot = 45° / 1.59 K / 13±1 mbar, MB = 62.6% |
| 1715 | Fill LN2 cold trap bath |
| 2125 | Start -80.3 V PuBe data series ## min long, SiHV-triggered, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10. Start series with the Start Button |
| 2125 | HV = -87.9 V, Vtest = -17.4 mV, Ileak = -19.2 uA, Vdet = -80.3 V (91%). |
| 2155 | Start -80.3 V PuBe data series 273 min long, NaI-triggered, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20. Start series with the Start Button |
| DAQ crash of some sort in the early AM. The 07220917_2155 directory has dumps up to 254 (2:13 am). There are more dumps in /midasraw/livedata, up to 272 (2:32 am). So if we can restart lazylogger those should get moved and renamed. | |
| UPDATE: got the DAQ working again a couple days later, now all 273 raw files are in the proper directory. |