start › cdms › k100 › run_diary › 20220919

2022/09/19 (Monday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on West side of lead/poly shield.
NaI trigger window: 7.6-11.2 MeV
Eighth 5K pumpout yesterday. Pumping IVC overnight, circulation and 1K Pot suspended.
0925 MC/1KPot/Sorb = 5.40/5.78/7.30 K. G1(fridge) = 540 mbar
0925 Turbo on IVC = 1.4 nbar. With Turbo valved out for 1 minute IVC = 12 nbar.
0925 Close IVC valves on top of fridge
0930 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (speedivalve closed).Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.24 bar on mechanical gauge.
0930 Pump out 1K Pot in stages, alternating 1K Pot Speedivalve and V4a, opening V4a in brief spurts to keep 4He pump pressure low (below 5 mbar).
0935 Start pumping 1K Pot continuously
0940 MC/1KPot/Sorb = 5.38/3.33/4.20 K.
1005 MC/1KPot/Sorb = 3.34/3.05/4.00 K. G1(fridge) = 412 mbar
1015 1K Pot = 0° / 2.93 K / 2.6 mbar
1015 Open 1K Pot needle valve → 15°. 1K Pot → 2.65 K / 6.5 mbar
1030 MC/1KPot = 3.07/2.43 K. G1(fridge) = 285 mbar
1030 Open 1K Pot needle valve → 30°. 1K Pot → 2.33 K / 8.0 mbar
1040 MC/1KPot = 2.89/2.12 K. G1(fridge) = 222 mbar
1040 Open 1K Pot needle valve → 45°. 1K Pot → 1.92 K / 13 mbar
1105 Connection to cdms1.spa.umn.edu is working again. Restart MIDAS. Looks like LazyLogger is copying leftover raw files from /midasraw/livedata to /midasraw/byseries, finishing up from the DAQ crash a couple days ago.
1110 MC = 2.18 K. G1(fridge) = 60 mbar. These are both flattening.
1115 1K Pot = 1.54 K / 7.5 mbar, kinda low. We'll see how it does when we start circulation
1120 MC drops relatively quickly from 2.16 to 2.07 K, I guess it pushed through the lambda point.
1155 MC/1KPot = 1.88/1.59 K. G1(fridge) = 45 mbar, still falling slowly
1155 1K Pot = 1.59 K / 13±1 mbar, this looks better.
1230 MC = 1.78 K. G1(fridge) = 37 mbar
1230 Close V3 and V9, leaving 35 mbar mixture in dump
1230 Turn on 3He pump
1235 Start circulation through Trap 1. V6 → 12.6%. Condenser peaks at 225 mbar. MC(H35) → 1.47 K. Still → 1.52 K / 50 mbar.
1240 Fill LN2 cold trap bath
1245 V6 → 13.0%. Condenser(G1) peaks at 240 mbar. MC(H35) → 1.32 K. Still(P1) → 1.35 K / 13 mbar.
1250 V6 → 13.4%. Condenser(G1) peaks at 230 mbar. MC(H35) → 1.10 K. Still(P1) → 1.20 K / 3.8 mbar.
1300 V6 → 13.8%. Condenser(G1) peaks at 196 mbar. MC(H35) → 0.55 K. Still(P1) → 1.17 K / 2.02 mbar.
1320 V6 → 14.4%. Condenser(G1) peaks at 203 mbar. MC(H35) → 0.40 K. Still(P1) → 1.15 K / 0.95 mbar.
1330 V6 → 15.0%. Condenser(G1) peaks at 199 mbar. MC(H35) → 0.20 K. Still(P1) → 1.17 K / 0.45 mbar.
1345 V6 → 100%. Condenser(G1) peaks at 192 mbar. MC(H35) → 96 mK. Still(P1) → 1.10 K / 0.13 mbar.
1420 MC = 30.0 mK. Condenser(G1) = 143 mbar. Still = 0.075 mbar.
1425 Open V14 to pull the rest of the mixture out of the dump into circulation. MC peaks at 60 mK, Still peaks at 26 mbar, Condenser peaks at 198 mbar, then all relax again. 1K Pot unaffected.
1450 Still = 0.6 mbar.
Close V14
1510 MC = 28.0 mK
1515 Plug in DCRCs 1&3
1520 Check phonon channels. They look good. Baselines still near -0.8 V except DCRC3pA - after a relock DCRC3pA is near -0.8 V too.
1605 MC = 29.1 mK. Seems to have settled here after turning on the DCRCs.
1700 Enable DCRC2 for HV control. But I forgot to set the ODB for 0V after manually dialing down the HV yesterday when the DAQ was inaccessible. So it jumped to -88V. Set the DCRC2qI back to 11.59 to get the HV back to nominally 0V. We'll let it settle.
1709 Dial HV up to -109.25V nominal, -107.2V actual, for 15% prebias (using RootSequencer)
1714 Dial HV down to -95V nominal, -92.9V actual
1717 HV = -92.9 V, Vtest = -23.4 mV, Ileak = -25.7 uA, Vdet = -82.6 V (89%). A little below 90% at the detector. Let's step it down a bit.
1718 Dial HV down to -90V nominal, -88.0V actual
1720 HV = -88 V, Vtest = -17.6 mV, Ileak = -19.3 uA, Vdet = -80.3 V (91%).
1723 Start -80.3 V PuBe data series 44 min long, SiHV-triggered, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
QET baselines seem more elevated than usual. Pulses look OK.
1730 MB = 69.0%
1808 Start -80.6 V PuBe data series 4 good min long, NaI-triggered, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20.
Start series with the Start Button
1810 HV = -88 V, Vtest = -16.9 mV, Ileak = -18.6 uA, Vdet = -80.6 V (92%).
TowerFE crash just 4 minutes into the data series
2012 Start -80.7 V PuBe data series 111 good min long, NaI-triggered, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20.
Start series with the Start Button
2015 HV = -88 V, Vtest = -16.5 mV, Ileak = -18.2 uA, Vdet = -80.7 V (92%).
Dump 111 is the last non-empty dump. Not sure what happened this time, some kind of crash.

Next Day