2022/09/16 (Friday)

Previous Day

  • Lab in-personnel: Zach
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on West side of lead/poly shield.
NaI trigger window: 7.6-11.2 MeV
Warmed up yesterday, seventh 5K pumpout.
Cooling to base today
1222 Turbo on IVC = 1.2 nbar. With Turbo valved out for 1 minute IVC = 12 nbar.
1224 Close IVC valves on top of fridge
1225 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (speedivalve closed).Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.23 bar on mechanical gauge.
1230 Pump out 1K Pot in stages, alternating 1K Pot Speedivalve and V4a, opening V4a in brief spurts to keep 4He pump pressure low (below 5 mbar).
1240 Start pumping 1K Pot continuously
1240 MC = 5.66 K, cooling slowly. 1K Pot = 3.37 K / 2.2 mbar, cooling slowly
1245 Open 1K Pot needle valve → 15°. 1K Pot → 3.06 K / 6.5 mbar
1248 Fill LN2 cold trap bath
1250 Open 1K Pot needle valve → 30°. 1K Pot → 2.93 K / 8.5 mbar
1255 Open 1K Pot needle valve → 45°. 1K Pot → 2.66 K / 13 mbar
1257 G1/G2 = 409/383 mbar dropping linearly
1300 MC/1Kpot = 3.25/2.43K dropping linearly
1424 Begin LHe transfer @ MB = 36.6%
1428 Close V3 and V9, leaving 49.5 mbar mixture in dump
1428 Turn on 3He pump
1435 Start circulation through Trap 1. V6 → 12.6%. Condenser peaks at 261 mbar. MC(H35) → 1.53 K. Still → 1.58 K / 50 mbar.
1445 V6 → 13.0%. Condenser(G1) peaks at 245 mbar. MC(H35) → 1.39 K. Still(P1) → 1.44 K / 20 mbar.
1449 P1 jumps for a moment, going over 30 mbar
1455 V6 → 13.4%. Condenser(G1) peaks at 221 mbar. MC(H35) → 1.19 K. Still(P1) → 1.20 K / 4.0 mbar.
1505 V6 → 13.8%. Condenser(G1) peaks at 188 mbar. MC(H35) → 0.99 K. Still(P1) → 1.13 K / 2.0 mbar.
1507 Ended LHe transfer @ MB = 94.5%
1513 G1 still rising slowly
1515 V6 → 14.4%. Condenser(G1) peaks at 200 mbar. MC(H35) → 0.63 K. Still(P1) → 1.12 K / 1.1 mbar.
1525 V6 → 15.0%. Condenser(G1) peaks at 190 mbar. MC(H35) → 0.30 K. Still(P1) → 1.14 K / 0.55 mbar.
1535 V6 → 100%. Condenser(G1) peaks at 182 mbar. MC(H35) → 0.14 K. Still(P1) → 1.08 K / 0.17 mbar.
1540 plug in DCRCs 1 and 3
1545 MC = 64mK, G1 = 166mbar
1650 MC = 28.4 mK and pretty flat, Condenser = 129 and pretty flat. 1K Pot = 1.57 K / 12±1 mbar.
1650 Open V14 to pull the rest of the mixture out of the dump into circulation.
MC rises to 85 mK, Still peaks above 30 mbar then falls again, Condenser rises to 208 mbar. 1K Pot unaffected.
as usual we'll wait till Still is below 1 mbar then close V14.
1715 Still = 0.8 mbar. close V14
1930 Swap HV leads to negative polarity
2030 Check phonon channels - three channels have shifted baselines. Relocking fixes two of them - had to adjust offset on DCRC1pA. Now all baselines close to -0.80 V
2042 Dial HV up to -109.25V nominal, -107.2V actual, for 15% prebias (using RootSequencer)
2047 Dial HV down to -95V nominal, -92.9V actual
2050 HV = -92.9 V, Vtest = -24.7 mV, Ileak = -27.2 uA, Vdet = -82 V (88%). A little worse than usual. Let's dial down to get Vdet up to 90%
2053 Dial HV down to -90V nominal, -88.0V actual
2055 HV = -88 V, Vtest = -19.5 mV, Ileak = -21.4 uA, Vdet = -79.4 V (90%).
2059 Start -79.4 V PuBe data series 59 min long, SiHV-triggered, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
2200 Start -79.7 V PuBe data series 747 min long, NaI-triggered, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20.
Start series with the Start Button
2205 HV = -88 V, Vtest = -18.9 mV, Ileak = -20.8 uA, Vdet = -79.7 V (91%).