start › cdms › k100 › run_diary › 20220915

2022/09/15 (Thursday)

Previous Day

  • Lab in-personnel: Zach
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on West side of lead/poly shield.
NaI trigger window: 7.6-11.2 MeV
Cooled to base on Monday, Sixth 5K pumpout.
Channels went normal overnight due to helium films, need to warm up
1355 MC/Coldplate = 31.3/150 mK, Still = 1.12 K / 0.075 mbar / 0 mW, G1/G2 = 132/128 mbar, 1K Pot = 45° / 1.59 K / 13±1 mbar, MB = 67.7%
1357 Fill LN2 cold trap dewar
I will begin another warm up and pump out
1400 Close V6 to stop circulation.
1405 V6 is now fully closed (light is off on IGH, Still pressure is rising, Condenser pressure falling). Turn off 3He pump
1405 Close 1K Pot needle valve
1410 Open V3 to let mixture from Condenser side into Still side.
1415 MC = 0.44 K
1415 Turn on Still and MC heaters to 20 mW each
1419 Turn off DAQ programs and unplug DCRCs 1&3
1425 Close 1K Pot speedivalve, V4a, 4He pump inlet valve. Turn off 4He pump. Close recovery valve to 4He pump outlet. NOTE: remember to open recovery valve when the pump is turned back on.
1500 MC/1KPot = 3.72/3.62
1500 G1(fridge) = 645 mbar. Open V9 to release mixture into dump
1545 MC temp starts to rise more quickly
1613 MC/1KPot = 5.70/4.35. Temperatures are relatively flat now.
1615 Check pressure in IVC: 0.10 ubar. Start pumping out IVC with turbo pump
1615 MC temp rising
1810 MC/Sorb= 10.1/10.2 K. Turn off MC and Still heaters