2022/08/30 (Tuesday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on West side of lead/poly shield.
Cooled to base yesterday after third 5K pumpout.
1135 Temperature spikes returned on schedule at around 4:50 am. For a while MC base was around 28.9 mK with spikes peaking mostly below 29.5 mK but occasional ones as high as 31 mK. Between 10 and 11 am there was a sort of wandering diversion up to around 32 mK then back down.
1140 MC/Coldplate = 29.5/150 mK, Still = 1.10K / 0.080 mbar / 0 mW, G1/G2 = 156/146 mbar, 1K Pot = 60° / 1.60 K / 15±1 mbar, MB = 70.3 %
1140 Phonon channels look good though some baselines are slightly lower than yesterday.
1400 Spent a couple of hours trying to get NaI triggers back reading on DCRC4. Enabling DCRC4 kept locking up MIDAS. Kept swapping to DCRC6 and trying different POE cables. Finally got it working in the original configuration. NOTE: signals are visible in DCRC4pA even when Polarity is set to 0.
1700 Count NaI triggers to estimate a rate. 3 pulses in 100 3.2768 ms traces → 9±5 Hz.
1700 MC base = 29.2 mK, with spikes.
1709 Dial HV up to -109.25V nominal, -107.3V actual, for 15% prebias (using RootSequencer)
1715 Dial HV down to -95V nominal, -93.0V actual
1720 Phonon channels look ok, although I had to reset offset on DCRC3pA
1720 HV = -93 V, Vtest = -17.5 mV, Ileak = -19.2 uA, Vdet = -85.3 V (92%).
NOTE: DCRC4 is enabled now. It hasn't been enabled in recent data. So we are back to recording NaI triggers.
1724 Start -85.3 V PuBe data series 60 min long, SiHV-triggered, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
1825 HV = -92.9 V, Vtest = -17.5 mV, Ileak = -19.3 uA, Vdet = -85.2 V (92%).
1829 Start -85.2 V PuBe data series 112 min long, NaI-triggered, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC3pA threshold 20.
Start series with the Start Button
2025 TowerFE crashed during Dump 113. Caught it pretty soon so not many empty dumps.
2030 HV = -92.9 V, Vtest = -18 mV, Ileak = -19.8 uA, Vdet = -85 V (91%).
2033 Start -85.0 V PuBe data series 44 min long, SiHV-triggered, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
2118 Start -84.8 V PuBe data series 723 min long, NaI-triggered, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC3pA threshold 20.
Start series with the Start Button
1920 HV = -92.9 V, Vtest = -18.4 mV, Ileak = -20.2 uA, Vdet = -84.8 V (91%).

next morning:

0935 TowerFE crash. Dumps after 723 are empty.
0935 HV = -92.9 V, Vtest = -20.3 mV, Ileak = -22.3 uA, Vdet = -84 V (90%).