UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20220704

2022/07/04 (Monday)

Previous Day

  • Lab in-personnel: Giulia
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe source behind lead wall.
NaI trigger window -t10760 -T15760 (7.6-11.2 MeV)
Taking NaI-triggered HV data overnight
0945 MC = 29.1 mK
0945 Start -197.3 V PuBe data series 12 min long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC3pD threshold 10 (Si-triggered).
Start series with the Start Button.
0950 Pulses still look fine.
0955 Having trouble stopping the MIDAS run, annoying. It's hanging on Stop Run and modifying the RunInfo stuff on the ODB isn't helping.
Eventually the Tower_FE and Trigger_FE programs crashed then I could stop the run. Restarted the programs and, luckily, I can still start a MIDAS run.
We'll do a test: take Si-triggered data while bringing HV back to 0V.
1000 Start -197.3 V to 0 V PuBe data series 30 min long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC3pD threshold 10 (Si-triggered).
Start series with the Start Button.
1004 Dial HV down to 0V. Rampdown starts at 10:04:17 and ends at 10:05:26. Then HV relaxes from -118 mV to -47 mV with a time constant around 2 minutes.
1140 Start 0 V PuBe data series 59 min long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC3pD threshold 10 (Si-triggered).
Start series with the Start Button.
1320 MC/Coldplate = 29.0/152 mK, Still = 1.08 K / 0.095 mbar / 0 mW, G1/G2 = 131/125 mbar, 1K Pot = 60° / 1.55 K / 8.0±0.5 mbar, MB = 66.1%
1330 Filled LN2 cold trap dewar
1331 Dial HV up to -230 V using RootSequencer for prebias
1345 Vtest = -340 mV, high.
If it was this high during last night's data, then the detector bias was much lower than -197.3 V - estimated V_detector only -48 V.
1355 Check again the purge gas flow at the Vent port on the LN2 Dewar. There was very little to no flow. Open valve just a bit more to restart flow
1357 Vtest = -100 mV, and still decreasing
1402 Vtest = -45 mV, better
1422 After increasing flow of purge gas a bit more, Vtest is down to about -10 mV. That means 98% of the HV is actually reaching the crystal.
1435 Relock DCRC3pD to get it back to standard baseline
1437 Start -197.3 V PuBe data series 6 min long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC3pD threshold 10 (Si-triggered).
Start series with the Start Button.
1444 Start -197.3 V PuBe data series 91 min long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20 (NaI-triggered).
Start series with the Start Button.
1445 MC = 29.2 mK
1617 Start -197.3 V PuBe data series 11 min long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC3pD threshold 10 (Si-triggered).
Start series with the Start Button.
1620 MC = 29.1 mK
1630 Start -197.3 V PuBe data series 4h31m long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20 (NaI-triggered).
Start series with the Start Button.
2102 Start -197.3 V PuBe data series 8 min long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC3pD threshold 10 (Si-triggered).
Start series with the Start Button.
2111 Start -197.3 V PuBe data series 552 dumps long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20 (NaI-triggered).
Start series with the Start Button.
2433 DAQ crash of some sort, after this point all dumps are empty. Only the first 221 dumps are good.

Next Day