start › cdms › k100 › run_diary › 20220316

2022/03/16 (Wednesday)

Previous Day

  • Lab in-personnel: Shubham, Matt, Giulia, Giovanna
Status Run 76 beginning. Warming up after LN2 transfer - IVC showed a leak during LN2 blowout. N2 gas in cryostat vacuum jacket. MC and Still heaters are on.
0835 MC(Cernox) = 222.0 K, warming at 8.9 K/hr. Assuming Troom = 72 F this implies a warming tau of 8.95 hours. That means the time to reach 0 C is 10.7 hours, which would be 7:16 pm tonight.
1035 Start pumping mixture out of fridge and cold LN2 Trap 1 into dump
1300 Close Condenser speedivalve and Still gate valve to isolate fridge
1315 Check contents of warm LN2 Trap 2, then pump out with Scroll Pump: 0.6→4.6→1.7 mbar.
1328 Start cryocleaning mixture through LN2 Cold Trap 2, V6 = 12.4%
Insert Trap2 into LN2 cylinder and fill the cylinder with LN2 upto brim. Check that He3 pump is ON. Use cryoclean process written note on the wall.
1339 Increase V6 → 12.5% to restart flow
1343 Increase V6 → 12.6% to restart flow
1352 Increase V6 → 12.7% to speed up flow
1433 Stop cleaning through Trap 2 after 65 min. G2 flow pressure was 580-760 mbar. Pump mixture to dump.
Removed 5.1(3.7) mbar.
1545 MC(Cernox) = 250 K, warming at 3.75 K/hr. Assuming Troom = 72 F this implies a warming tau of 12.1 hours. That means the time to reach 0 C is 8.6 hours, which would be a little after midnight tonight.
1655 Check contents of warm LN2 Trap 1, then pump out with Scroll Pump: 0.1→6.5→1.3 mbar. This trap was used during the LN2 cooldown.
1705 Check pressure in IVC, it's a little low for the current temperature. Let more N2 into IVC to about 0.87 atm. MC(Cernox) reads 253.5 K now so should rise to about 1.01 atm at room temperature.
1715 Check pressure in cryostat vacuum jacket: 25 mbar. (MC(Cernox) = 254 K).
1918 Start pumping out cryostat vacuum jacket: 1) check that cryostat vacuum valve is open 2) plug in scroll pump 3) open scroll pump valve 4) wait till pressure is below 0.1 mbar 5) open turbo pump valve (just above Pfeiffer turbo pump) 6) close scroll pump valve 7) unplug scroll pump. Leave it turbo-pumping overnight