| Status | Run 76 prep. All mixture in dump. LHe Trap and LN2 Trap 1 still have contaminants. Cryostat Vacuum Jacket pumped out. Tower removed. |
| 1105 | Tape wires from 2D source mover Y-stripline down on bench, use very fine sandpaper to scrape off insulation. Turn stripline over and repeat. Ohmmeter shows ~ 90 Ohm for each coil. |
| 1110 | Trim 2D source mover Y-stripline wire ends to around 1.5 times the length of the solder post on the connectors |
| 1125 | Begin soldering Y-stripline wires to mini connectors. Twist wire around post with tweezers. Wound up re-sandpapering with somewhat less fine sandpaper to get a better surface for solder to stick. |
| 1145 | Finish soldering connector to Y-stripline. Connector is at a right angle to stripline, should help to plug in to tower. |
| 1155 | Sandpaper ends of wires on X-stripline. With the coarser grit I accidentally broke one wire kinda short. Will still work, but need to be careful. |
| 1215 | Finish soldering connector to X-stripline. |
| 1220 | Cover connector base on 2D source mover striplines with Kapton tape strip on each side. |
| 1225 | Connect 2D source mover stiplines to controller connectors. |
| 1230 | Connect 2D source mover to controller, plug in controller, turn on controller software on laptop. Try some programs, seems to move properly. |
| 1330 | More bench tests of 2D source mover. Turned the mover upside down, supported on two wire spools - all basic functions (including piezo recalibration) seem to work. |