start › cdms › k100 › run_diary › 20220131

2022/01/31 (Monday)

Previous Day

  • Lab in-personnel: Matt
Status Run 75 in progress
Setup Tower Faces 5/3/2 → Cables 2/1/3 → nothing/diverter_board/nothing → DCRC55(1)/DCRC50(3)/DCRCx(4)
Lead shield in place. Am Source y = 0.03 mm
1335 Goldman/Tower/MC(H35)/Coldplate = 41.8/72.8/34.3/185 mK, Still = 1.43 K / 0.110 mbar / 0 mW, G1/G2= 264/251 mbar, 1K Pot = 1.59 K / 11±1 mbar, MB = 64.1%.
1340 Fill LN2 cold trap bath.
1435 Check all the pins that should be connected to bias on the Side1-readout, Side2-HV board. Two of them seem to not be connected well (R36 and R42).
1450 Repair solder joints on R36 and R42 on the Side1-readout, Side-2 HV board.

Current phonon settings shown below. We'll use the same Lockpoint and SQUIDbias for each detector channel after switching to HV. And possibly the same VGAGain values for each DCRC.

[/Equipment/Tower01/Settings/DCRC1/Phonon]
VGAGain (multiplicative factor) = FLOAT[4] :
[0] 5
[1] 5
[2] 2
[3] 5
ADCOffset (V) = FLOAT[4] :
[0] 0.4
[1] 0.2
[2] 0.2
[3] 0
LockpointAdjust (mV) = FLOAT[4] :
[0] -3.5
[1] -3.8
[2] -4
[3] -2.5
SQUIDBias (uA) = FLOAT[4] :
[0] -80
[1] -95
[2] -91
[3] -102
QETBias (uA) = FLOAT[4] :
[0] 239.9
[1] 283.5
[2] 275.3
[3] 199.9
[/Equipment/Tower01/Settings/DCRC3/Phonon]
VGAGain (multiplicative factor) = FLOAT[4] :
[0] 5
[1] 5
[2] 5
[3] 5
ADCOffset (V) = FLOAT[4] :
[0] 0.4
[1] 0.3
[2] 0.2
[3] 0.5
LockpointAdjust (mV) = FLOAT[4] :
[0] -3
[1] -3.4
[2] -2.9
[3] -3.4
SQUIDBias (uA) = FLOAT[4] :
[0] -50
[1] -70
[2] -70
[3] -60
QETBias (uA) = FLOAT[4] :
[0] 180.6
[1] 174.6
[2] 220.1
[3] 0
[/Equipment/Tower01/Settings/DCRC4/Phonon]
VGAGain (multiplicative factor) = FLOAT[4] :
[0] 2
[1] 3
[2] 3
[3] 3
ADCOffset (V) = FLOAT[4] :
[0] 0.3
[1] 0.3
[2] 0.2
[3] 0.9
LockpointAdjust (mV) = FLOAT[4] :
[0] -4
[1] -5.4
[2] -2.3
[3] -5.7
SQUIDBias (uA) = FLOAT[4] :
[0] -70
[1] -80
[2] -70
[3] -75
QETBias (uA) = FLOAT[4] :
[0] 230.7
[1] 164.4
[2] 258.9
[3] 202.5

Next Day