2021/12/31 (Friday)

Previous Day

  • Lab in-personnel: Matt
Status Run 75 in progress
Setup Tower Faces 5/3/2 → Cables 2/1/3 → nothing/diverter_board/nothing → DCRC55(1)/DCRC50(3)/DCRCx(4)
Lead shield in place. Am Source y = 2.07 -2.02 mm
1600 Goldman/Tower/MC(H35)/Coldplate = 47.6/84.5/37.3/183 mK, Still = 1.26±0.02 K / 0.120 mbar / 0 mW, G1/G2= 242/229 mbar, 1K Pot = 1.62 K / 19±1 mbar, MB = 27.1%.
Coldplate and Still thermometers seem to have fallen into their low states
1605 Do mini-IbIs on all channels, find QET bias corresponding to dP/dI = 1.0 uV
Goldman temperature 50.85 mK 49.95 mK 47.60 mK
QET bias (uA)
Channel Previous Previous New
DCRC1pA 241 (0.94) 244 (0.89) 248.7
DCRC1pB 314 317 323
DCRC1pC 274 (0.95) 277 (0.84) 284.5
DCRC1pD 178 (0.79) 181 (0.59) 190.5
DCRC3pA 160 (0.91) 161 (0.59) 169.6
DCRC3pB 162 (1.04) 162 (0.66) 170.5
DCRC3pC 217 (0.98) 217 (0.94) 221.9
DCRC3pD - - -
DCRC4pA 205 (0.90) 209 (0.90) 212.4
DCRC4pB 148 (0.93) 149 (0.81) 153.0
DCRC4pC 256 (0.97) 258 (0.91) 262.8
DCRC4pD 189 (0.90) 190 (0.73) 198.7

Note: number in parentheses is the dP/dI for that setting based on the following IbIs measurement

1717 Begin LHe transfer @ MB = 24.6%
1720 Fill LN2 cold trap bath
1748 End LHe transfer @ MB = 90.3%