start › cdms › k100 › run_diary › 20211230

2021/12/30 (Thursday)

Previous Day

  • Lab in-personnel: Matt
Status Run 75 in progress
Setup Tower Faces 5/3/2 → Cables 2/1/3 → nothing/diverter_board/nothing → DCRC55(1)/DCRC50(3)/DCRCx(4)
Lead shield in place. Am Source y = 2.07 -2.02 mm
1505 Goldman/Tower/MC(H35)/Coldplate = 48.4/85.4/37.7/214 mK, Still = 1.70±0.05 K / 0.120 mbar / 0 mW, G1/G2= 242/237 mbar, 1K Pot = 1.62 K / 19±1 mbar, MB = 36.4%.
1510 Mini-IbIs on DCRC1pA: new QET bias for 1uV dP/dI is 248.2 uA
1540 Screwy behavior with DCRC1pC IbIs - the normal part at high QET bias is not linear, it has jumps. Dunno what's wrong. Let's zap SQUIDs, maybe that will help?
1540 Zap SQUIDs on all 3 DCRCs. Surprisingly strong temperature response, Goldman peaks at 76 mK.
1545 Fill LN2 cold trap bath
After zap DCRC1pC offset seems to be in a more normal range.