2021/12/28 (Tuesday)

Previous Day

  • Lab in-personnel: Matt
Status Run 75 in progress
Setup Tower Faces 5/3/2 → Cables 2/1/3 → nothing/diverter_board/nothing → DCRC55(1)/DCRC50(3)/DCRCx(4)
Lead shield in place
1130 Goldman/Tower/MC(H35)/Coldplate = 49.4/87.2/38.6/215 mK, Still = 1.70±0.04 K / 0.130 mbar / 0 mW, G1/G2= 238/234 mbar, 1K Pot = 1.63 K / 19±1 mbar, MB = 69.3%.
1140 Fill LN2 cold trap bath
1210 Goldman = 50.0 mK
1215 Start some mini-IbIs, find QET biases for dP/dI = 1.0 uV
Goldman temperature 50.85 mK 49.95 mK
QET bias (uA)
Channel Previous New
DCRC1pA 241 (0.94) 244
DCRC1pB 314 317
DCRC1pC 274 (0.95) 277
DCRC1pD 178 (0.79) 181
DCRC3pA 160 (0.91) 161
DCRC3pB 162 (1.04) 162
DCRC3pC 217 (0.98) 217
DCRC3pD - -
DCRC4pA 205 (0.90) 209
DCRC4pB 148 (0.93) 149
DCRC4pC 256 (0.97) 258
DCRC4pD 189 (0.90) 190

The number in parentheses indicates what the dP/dI in uV would be today at yesterday's QET bias. So a 0.9 mK change in temperature results in an average 7% change in dP/dI.

1350 Start 0V data series, 35 min long, with 1000 initial randoms, no in-run randoms, trigger on DCRC3pB threshold 5.
Start series with the Start Button.

Next Day