2021/12/27 (Monday)

Previous Day

  • Lab in-personnel: Matt
Status Run 75 in progress
Setup Tower Faces 5/3/2 → Cables 2/1/3 → nothing/diverter_board/nothing → DCRC55(1)/DCRC50(3)/DCRCx(4)
Lead shield in place
1105 Goldman/Tower/MC(H35)/Coldplate = 50.0/121/37.7/213 mK, Still = 1.69±0.06 K / 0.130 mbar / 0 mW, G1/G2= 231/228 mbar, 1K Pot = 1.63 K / 19±1 mbar, MB = 23.6%.
1110 Fill LN2 cold trap bath
1300 Tried applying some Still heat. Goldman thermometer was at 50.8 mK after LN2 fill.
Still = 0 mW : 49.8 mK.
Still = 3 mW → 49.8 mK.
Still = 6 mW → 49.6 mK.
Still = 0 mW → 50.7 mK.
1355 Perform mini IbIs on DCRC1pA to find QET bias corresponding to dP/dI = 1 uV: 241 uA.
1505 Finish mini-IbIs on all channels, find QET biases for dP/dI = 1 uV on each: see below:
Goldman temperature 50.85 mK
Channel QET bias (uA)
DCRC1pA 241
DCRC1pB 314
DCRC1pC 274
DCRC1pD 178
DCRC3pA 160
DCRC3pB 162
DCRC3pC 217
DCRC3pD -
DCRC4pA 205
DCRC4pB 148
DCRC4pC 256
DCRC4pD 189

note: DCRC1pB has a weird normal part of the IbIs curve so it's hard to do mini-IbIs. QET bias above is a guess based on how each QET bias changed from that calculated from IbIs taken a few weeks ago.

1522 Begin LHe transfer @ MB = 14.9%
1558 End LHe transfer @ MB = 93.6%

Next Day