| Status | Run 75 prep. |
| | All mixture is in dump. Fridge isolated. 1K Pot isolated. LN2 Trap 2 still has contaminants in it. MB in pit, clean tent erected. Cryostat vacuum jacket (CVJ) pumped out. IVC open. Tower removed. |
| | 100 mK stage and 600 mK stage installed yesterday. |
| 1330 | Bring SQUET ZC-70 to wirebonder lab for modifications: reduce shunt on all four phonon channels to 6.7 mΩ. Wirebonder power 475/350, bond time 40/40. Scraped copper with hypodermic first. |
| 1330 | Tower installation: install 4K stage |
| 1430 | Begin installing SQUETs, choosing which stripline goes to which Tower face. |
| 1445 | Electrical check of all phonon-related lines on ZC-70, all good. |
| 1530 | Spent a long time straightening pins on interposer so that they would plug into a stripline. Used another interposer board to straighten the pins one row at a time, finally plugged the interposer into one of the old striplines on the bench. Then it was able to plug easily into a stripline. |
| 1630 | At Tower Face 5/2/3 plugged in SQUETs ZC-13/ZC-70/ZC-39+interposer plugged into Striplines 41/43/48 |