Previous Day

  • Lab in-personnel: Giulia and Matt
Status Run 74 in progress. First cooldown after IVC rebuild.
Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/Source_mover_board+HV_diverter_board → DCRC50(3)/55(1). Kepco 40V supplye → HV BNC. No lead/poly shield. Currently both DCRCs are on.
Position of Am source: y = -5.01 mm (mostly over blocker)
now we'll move the source back to the center (doing this from home)
0851 Do mvNwait +4steps×76 times, 40 second wait, with dwell = 40 and power = 6.8 V. (Around 5.93 mm/hour)
This sequence will take 50min40sec and will move the source a total distance of 5.01 mm. Should end at 9:41 pm.
Mean Tower temperature settles around 1.71 K during move.
0940 Tower/MC/Coldplate = 1821/176.4/242 mK, Still = 1.35 K / 0.095 mbar / 0 mW, G1/G2= 324.1/296.7 mbar, 1K Pot = 1.59 K / 20±1 mbar. MB = 35.2%.
The values above were recorded while the source was moving.
0952 Fill LN2 cold bath.
1115 At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 1.7/1.1/1.3/5.0 8e-6. This is q3/q1/q4/q2
At 3kHz (peak) noise in DCRC3qI/qO/DCRC1qI/qO = 8.5/1.9/2.0/16 e-6.
Noise is considerably lower than yesterday.
1210 Tower/MC = 121/36.0 mK.
1215 Turn on MC heater to 300 μW. We'll try taking data at an elevated temperature again (last time was July 23
1245 MC = 436 mK and flat. Tower = 404 mK and flat.
1305 Start 50 min series of 0V/-10V (bias applied after 60 sec), at elevated temperature, with 1000 beginning randoms, trigger threshold 60/50/70/100 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Start sequence with Start button.
07210818_1305. No difference between this 0.4 K data and base temp data
1400 Dial bias to 0V.
1400 Turn off MC heater.
1410 Flash detector for 10 minutes with DCRC1&3 LEDs.
1545 Erect lead shield around fridge. 4 walls (no top pieces).
1635 Tower/MC = 92/35.7 mK
1634 Start 45 min series of 0V/-10V (bias applied after 60 sec), with shield present, with 1000 beginning randoms, trigger threshold 50/40/40/60 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Start sequence with Start button.
1719 Dial HV down to 0V and flash for 10 minutes with DCRC1&3 LEDs.
07210818_1634
1658 Begin LHe transfer @ MB = 24.2 %
1731 End LHe transfer @ MB = 91.3 %

Next Day