2021/08/17 (Tuesday)
- Lab in-personnel: Giulia and Matt
| Status | Run 74 in progress. First cooldown after IVC rebuild. |
|---|---|
| Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels. | |
| Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/Source_mover_board+HV_diverter_board → DCRC50(3)/55(1). Kepco 40V supplye → HV BNC. No lead/poly shield. Currently both DCRCs are on. | |
| Position of Am source: y = 0.00 mm | |
| 0901 | Tower/MC/Coldplate = 202.8/36.1/192.5 mK, Still = 1.39 K / 0.075 mbar / 0 mW, G1/G2= 292.6/270.2 mbar, 1K Pot = 1.59 K / 20±1 mbar. MB = 70.8%. |
| 0909 | Fill LN2 cold bath. |
| 0910 | At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 35/4.7/3.7/90 8e-6. This is q3/q1/q4/q2 At 3kHz (peak) noise in DCRC3qI/qO/DCRC1qI/qO = 940/92/74/1100e-6 |
| Noise is considerably higher than yesterday. | |
| 1043 | Start data then abort after randoms, overtriggering |
| 1047 | Start 45 min series of 0V/-12V (bias applied after 60 sec), with 1000 beginning randoms, trigger threshold 120/60/65/300 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Start sequence with Start button. |
| 1135 | Dial HV down to 0V and flash for 10 minutes with DCRC1&3 LEDs. |
|
| 07210817_1047. Detector is still in a strong impact ionization “state”. This peaks even higher than 12V did on Friday and yet 9V is weak. |
| we'll try 0V/-10V | |
| 1215 | Start 47 min series of 0V/-10V (bias applied after 60 sec), with 1000 beginning randoms, trigger threshold 120/50/70/280 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Start sequence with Start button. |
| 1304 | Dial bias down to 0V and flash for 10 minutes with DCRC1&3 LEDs. |
| Let's move near the blocker, to see if a lower rate of Am events changes the impact ionization pattern. | |
| 1329 | Do mvNwait -4steps×69 times, 40 second wait, with dwell = 40 and power = 6.8 V. (Around 5.93 mm/hour) This sequence will take 46min0sec and will move the source a total distance of 4.55 mm. Should end at 2:15 pm. Mean Tower temperature settles around 1.80 K during move. |
| 1415 | Move sequence ends, pretty much on time. |
| Position of Am source: y = -4.55 mm. The edge of the blocker should be at -4.59 mm. I expect this to block roughly 50% of the Am-241 events. | |
| 1527 | Start 45 min series of 0V/-10V (bias applied after 60 sec), Am source y= -4.55 mm, with 1000 beginning randoms, trigger threshold 50/40/60/90 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Start sequence with Start button. |
| 1612 | Dial HV down to 0V and flash for 10 minutes with DCRC1&3 LEDs. |
|
| 07210817_1527. Looks like with reduced 60keV rate, the impact ionization is pretty much the same although the falloff after the peak seems slower |
| Let's move further over the blocker. We'll move about 1 sigma of the source spread, which is around 0.47 mm according to sim. | |
| 1642 | Do mvNwait -4steps×7 times, 40 second wait, with dwell = 40 and power = 6.8 V. (Around 5.93 mm/hour) This sequence will take 4min40sec and will move the source a total distance of 0.46 mm. Mean Tower temperature peaks around 1.70 K during move. |
| 1647 | Move sequence ends. |
| 1747 | Start 45 min series of 0V/-10V (bias applied after 60 sec), with 1000 beginning randoms, trigger threshold 70/80/70/90 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Start sequence with Start button. |
| 1832 | Dial HV down to 0V and flash for 10 minutes with DCRC1&3 LEDs. |