start › cdms › k100 › run_diary › 20210730

2021/07/30 (Friday)

Previous Day

  • Lab in-personnel: Giulia
Status Run 74 in progress. First cooldown aftr IVC rebuild.
Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/Source_mover_board+HV_diverter_board → DCRC50(3)/55(1). Kepco 40V supplye → HV BNC. No lead/poly shield. Currently both DCRCs are on.
1008 Tower/MC/Coldplate = 111.2/33.7/ 193.5 mK, Still = 1.42 K / 0.075 mbar / 0 mW, G1/G2= 277.6/247.2 mbar, 1K Pot = 1.61 K / 20±1 mbar. MB = 70.1%.
1021 Fill LN2 cold trap bath.
1200 Problem with /data/cdms1, which is where we store raw data - currently the computers can't mount this disk. We are able to start MIDAS and view pulses, but we will not be able to write data until this is fixed.
1245 At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 12/1.6/2.3/36 e-6. This is q3/q1/q4/q2
At 3kHz (peak) noise in DCRC3qI/qO/DCRC1qI/qO = 380/39/38/940 e-6
These are all better than yesterday. 3kHz oscillations are still pretty significant though.