UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20210731

2021/07/31 (Saturday)

Previous Day

  • Lab in-personnel: Giulia and Zach
Status Run 74 in progress. First cooldown aftr IVC rebuild.
Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/Source_mover_board+HV_diverter_board → DCRC50(3)/55(1). Kepco 40V supplye → HV BNC. No lead/poly shield. Currently both DCRCs are on.
1148 Tower/MC/Coldplate = 100.7/33.4/ 190.6 mK, Still = 1.40 K / 0.075 mbar / 0 mW, G1/G2= 271.3/248.6 mbar, 1K Pot = 1.61 K / 20±1 mbar. MB = 32.8%.
1201 Fill LN2 cold trap bath.
1230 Start LHe transfer @ MB = 32.2 %
1300 End LHe transfer @ MB = 87.4%
1320 Zach and Giulia leave lab.

Next Day