2021/07/29 (Thursday)
- Lab in-personnel: Giulia and Pramod
| Status | Run 74 in progress. First cooldown aftr IVC rebuild. |
|---|---|
| Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels. | |
| Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/Source_mover_board+HV_diverter_board → DCRC50(3)/55(1). Kepco 40V supplye → HV BNC. No lead/poly shield. Currently both DCRCs are on. | |
| 1024 | Tower/MC/Coldplate = 97.3/33.8/ 192.8 mK, Still = 1.40 K / 0.08 mbar / 0 mW, G1/G2= 275.2/251.8 mbar, 1K Pot = 1.61 K / 20±1 mbar. MB = 31.6%. |
| 1132 | Fill LN2 cold trap bath. |
| 1404 | At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 13/2.5/2.3/51 e-6. This is q3/q1/q4/q2 At 3kHz (peak) noise in DCRC3qI/qO/DCRC1qI/qO = 575/61/55/1584 e-6 |
| These are similar to yesterday. | |
| 1431 | Start 45 min series of 0V/-15V (bias applied after 60 sec), Tower @ 103 mK, with 1000 beginning randoms, trigger threshold 80/70/60/70 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Start sequence with Start button. |
| 1527 | Dial bias down to 0V and flash detector for 600 seconds |
| 1551 | Begin LHe Transfer, MB = 22.3%. |