2021/07/13 (Tuesday)

Previous Day

  • Lab in-personnel: Pramod,Giulia
Status Run 74 in progress. First cooldown after IVC rebuild.
Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/Source_mover_board+HV_diverter_board → DCRC50(3)/55(1). Kepco 40V supplye → HV BNC. No lead/poly shield. Currently DCRC55(1) is disconnected and off, Cable 3 plugged into source mover board only.
0954 Tower/MC/Coldplate = 155/34.5/185 mK, Still = 1.29 K / 0.075 mbar / 0 mW, G1/G2= 286.9/262.6 mbar, 1K Pot = 1.63 K / 20±1 mbar. MB = 33.3%
1002 Fill LN2 cold trap bath.
1145 Get MIDAS working again. There have been some IT changes that made things not work… currently we can start MIDAS (on cdms1) but we can only see the webpage from a browser run on cdms1, not on other computers in the cluster.
1145 At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 23/3.7/2.5/47 e-6. This is q3/q1/q4/q2
At 3kHz (peak) noise in DCRC3qI/qO/DCRC1qI/qO = 1400/150/110/2300 e-6
3kHz noise is strong again. One possibility - it's pretty humid today, and the LN2 E-box purge may have been off overnight. Turned the purge on just now at a pretty high rate.
1203 Start 84 min series of 0V/-15V (bias applied after 200 sec, bias removed after 66 min), with 1000 beginning randoms, trigger threshold 150/60/30/300 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Start sequence with Start button.
1343 Begin LHe transfer @ MB = 28.2%
1414 End LHe transfer @ MB = 90.1%