2021/07/14 (Wednesday)

Previous Day

  • Lab in-personnel:Pramod and Giulia
Status Run 74 in progress. First cooldown after IVC rebuild.
Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/Source_mover_board+HV_diverter_board → DCRC50(3)/55(1). Kepco 40V supplye → HV BNC. No lead/poly shield. Currently DCRC55(1) is disconnected and off, Cable 3 plugged into source mover board only.
1051 Tower/MC/Coldplate = 195/36.3/171 mK, Still = 1.39 K / 0.08 mbar / 0 mW, G1/G2= 295.9/271.5 mbar, 1K Pot = 1.62 K / 20±1 mbar. MB = 65%
1059 Fill LN2 cold trap bath.
1710 At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 11/2.3/1.6/11 e-6. This is q3/q1/q4/q2
At 3kHz (peak) noise in DCRC3qI/qO/DCRC1qI/qO = 905/75/48/715 e-6
These are all lower than yesterday. Maybe because we have the E-box purge going? Yesterday morning it was off.
1734 Start 45 min series of 0V/-15V (bias applied after 2.5 min), with 1000 beginning randoms, trigger threshold 90/50/35/90 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Start sequence with Start button.
1838 The voltage across the test resistor, V_test =1.4 mV.
From today's data series
here's a series from March with the same bias