start › cdms › k100 › run_diary › 20210521

2021/05/21 (Friday)

Previous Day

  • Lab in-person: Matt
  • Virtual partner: Zach
Status Run 74 in progress. Warmed up after 3rd appearance of IVC leak at 4K. Cryostat Vacuum Jacket pumped out. Beginning reassembly of IVC.
Yesterday the stripline tube was installed.
1200 Matt enters lab
1210 Clean all three indium surfaces on upper IVC section one more time
1235 Put all pieces of the nine fridge poly layers back in place and secure with Kapton tape.
I can see a longitudinal scratch on the outside of the stripline tube, facing out and a little to the right. This is concerning, and I'm wondering whether I should remove it again, polish it again, and identify where the scratch occurs. This must be at one of the fridge baffles.
1250 Pull out stripline tube again. Actually there don't seem to be new scratches at the business end, at least not big ones. You can hear scraping as the tube is moved down, but it seems like it could be from the stripline baffles inside rather than the fridge baffles.
1255 Polish business end of stripline tube again, with two different kinds of Scotch Brite then finish with 1500-grade sandpaper. Looks shiny.
1300 Put two stripes on end of stripline tube with red Sharpie, one at the edge of the polished section (there's a noticeable cut in places from when Bill V turned it in the lathe a few years ago) and another one 1 cm higher. We can look for new scratches at the end.
1320 Put another layer of Kapton tape over the edges of all the fridge baffle holes, just one piece centered a little right of directly outward, since this seems to be where the longitudinal scratch occurred (plus zigzag scratches from past installations).
1330 Tape foil clumps between plates in bottom four stripline baffles
1345 Raise stripline tube up through baffles again. Some new zigzag scratches are evident. We'll take it out again and consider more seriously sanding those baffle edges.
1255 Remove stripline tube again, polish out the scratches again
1530 With some small files I borrowed from Mike Rother, file the inner surface of the bottom fridge baffle, then sand it with 180 then 1500 -grit sandpaper, then cover edge with Teflon tape (don't know if that will be any better than Kapton)
1600 Do the same with the second fridge baffle.
1630 Do the same with the third fridge baffle. Two more to go.
1645 Matt exits lab