start › cdms › k100 › run_diary › 20210405

2021/04/05 (Monday)

Previous Day

  • Lab in-person: Matt
  • Virtual partner: Zach
Status Run 74 in progress. Warmed the fridge to fix short and thermal touch. Helium gas in cryostat vacuum jacket, partially pumped out. Mixture cryocleaned four times. IVC sealed and pumping for several days.
We've been seeing a high leak signal in the IVC, slowly falling as we continue to pump.
1025 Matt enters lab
1030 Turbo on IVC = 16 nbar. With Turbo valved out for 1 minute, IVC = 250 nbar.
1055 Pump on IVC with LD, close Turbo valve and wait 1 minute: 162 pbarL/s.
It's continuing to improve, slowly. It seems to be falling exponentially. At the current trend, tomorrow it should fall below 100 pbarL/s, and it won't reach 10 pbarL/s until the 10th, 5 days from now.
1105 Start cryocleaning mixture through LN2 Cold Trap 1, V6 = 12.4%
1125 Increase V6 → 12.6% to restart flow
1145 Increase V6 → 12.7% to speed up flow
1205 Stop cleaning through Trap 1 after 60 min. G2 flow pressure was 585-700 mbar. Pump mixture to dump.
Removed 4.3 mbar.
1125 Leak signal settles around 189 pbarL/s. Squirt around IVC joint - no increase. Squirt around stripline elbow: signal rises to ~500 pbarL/s. Ok so that leak is still present.
1135 Open Turbo Pump valve a little, enough so that the leak signal falls below 100 pbarL/s, so that maybe it will be more sensitive. Around 93.0 pbarL/s, fluctuating by a few 10ths. Squirt around IVC joint: leak < 0.5 pbarL/s. There may have been an increase of a few 10ths but that also may have been helium floating up to the stripline elbow. I think the IVC seal is OK.
1150 Remove clean tent from bottom of fridge. Reattach fill tube with funnel to IVC exterior.
1200 Removed covers from pit. Used the long 2×4 to put through the handles to prevent covers from falling in. Wore gloves to avoid splinters.
1235 Start cryocleaning mixture through LN2 Cold Trap 2, V6 = 12.8%
1245 Increase V6 → 12.9% to speed up flow
1315 Increase V6 → 13.0% to speed up flow
1345 Stop cleaning through Trap 2 after 70 min. G2 flow pressure was 585-700 mbar. Pump mixture to dump.
Removed 3.7 mbar.
1250 Finish raising Main Bath and replace pit covers, again using the long 2×4
1435 Finish tightening all Main Bath bolts. Note that several of the small bolts required 2 washers. Probably makes sense to use 2 washers on all of them next time.
1435 Pump on IVC with LD, close Turbo valve and wait 1 minute: 155 pbarL/s.
1440 Check contents of warm LN2 Trap 1: 4.3 mbar. Pump out with Scroll Pump
1500 Apply duct tape around stripline elbow in hopes of blocking some of the helium that might leak in during transfers
1530 Matt exits lab