UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20210228

2021/02/28 (Sunday)

Previous Day

  • Lab in-person: Ruiren
  • Virtual partner: Matt
Status Run 74 in progress.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/motor_board+HV_board → DCRC50(3)/55(1). No lead/poly shield.
0830 MB = 70.7 %
0850 Tower = 94.1 mK
0850 Start 31min series of 3V data (bias applied after 1 min), with 1000 beginning randoms, trigger threshold 50/50/50/50 on DCRC3qI/qO/DCRC1qI/qO, 5 Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Use sequence r74_takeandflash_dcrcbias.msl
Looks like it may have been triggering constantly. Annoying.
1125 Tower = 94.9 mK
1125 Start 31min series of 1V data (bias applied after 1 min), with 1000 beginning randoms, trigger threshold 50/50/50/50 on DCRC3qI/qO/DCRC1qI/qO, 5 Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Use sequence r74_takeandflash_dcrcbias.msl
Triggering looks OK.
1320 Tower = 96.4 mK
We'll try 3V again, maybe this time it won't overtrigger.
1319 Start 31min series of 3V data (bias applied after 1 min), with 1000 beginning randoms, trigger threshold 50/50/50/50 on DCRC3qI/qO/DCRC1qI/qO, 5 Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Use sequence r74_takeandflash_dcrcbias.msl
Triggering looks ok this time
1608 Ruiren enters lab
1612 fill lN2 cold trap bath
1612 Tower/MC/Coldplate = 96.7/31.3/187 mK, Still = 1.49 K / 0.130 mbar / 0 mW, G1/G2= 189/180 mbar, 1K Pot = 1.67 K / 36±2 mbar.
1614 Ruiren leaves lab
2343 Start 31min series of 7V data (bias applied after 1 min), with 1000 beginning randoms, trigger threshold 40/40/40/40 on DCRC3qI/qO/DCRC1qI/qO, 5 Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Use sequence r74_takeandflash_dcrcbias.msl
Triggering looks OK