| Status | Run 74 in progress. |
| | Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/motor_board+HV_board → DCRC50(3)/55(1). No lead/poly shield. |
| 1500 | Matt enters lab |
| 1500 | Tower/MC/Coldplate = 121/31.2/184 mK, Still = 1.43 K / 0.130 mbar / 0 mW, G1/G2= 78/74 mbar, 1K Pot = 1.72 K / 80±10 mbar. |
| 1505 | MB = 56.7% |
| 1510 | Fill LN2 cold trap bath |
| 1515 | Take a look at the charge channels. The traces in DCRC1 seem to have wandering baselines - this was seen in the data yesterday too. Should try processing that data to quantify this. Tried applying 5V to all channels - saw occasional pulses in DCRC3 channels, not in DCRC1 channels. |
| 1520 | Matt exits lab |