  • Lab in-person: Matt
  • Virtual partner: Derek
Status Run 74. Started circulation yesterday.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/motor_board+HV_board → DCRC50(3)/55(1). No lead/poly shield.
0955 MC(H35) = 31.2 mK, Still = 0.140 mbar, G1/G2 = 78/73 mbar. 1K Pot = 40° / 1.72 K / 80±10 mbar
1040 Looked at the channels a bit with the Pulse Viewer. Can see noise on all channels. Tried biasing and setting a trigger - didn't really see pulses. MC temperature rose a little.
let's try a flash, see if the LEDs work.
1040 Flash LEDs on DIBs 1&3 for 30 seconds. MC peaks at 63 mK. That's not very high.
1055 Try turning on all 4 LEDs and leaving them on. Note - current is set to 0.6 mA. Temperature is rising, though not super-fast. I guess I'm used to looking at Tower though, not MC. After 5 minutes MC → 150 mK.
1100 Turn off LEDs.
1135 MC(H35) = 31.7 mK, within 0.5 mK of base.
1200 MC(H35) = 31.3 mK, within 0.1 mK of base.
1315 Matt enters lab
1320 Tower/MC/Coldplate = 126/31.2/187 mK, Still = 1.51 K / 0.140 mbar / 0 mW, G1/G2= 78/74 mbar, 1K Pot = 1.72 K / 80±10 mbar.
Tower is surprisingly warm compared to last run. Coldplate and Still are also significantly warmer.
1325 MB = 20.6%
1355 Begin LHe Transfer, MB = 20.0%
1405 Fill LN2 cold trap bath
1431 Finish LHe Transfer, MB = 89.8%
1450 1K Pot temperature and pressure fell a bit during the transfer. Tower temperature is rising after the transfer, now at 200 mK and climbing. Reminiscent of Run 73. Luckily it doesn't matter as much for this run.
1450 Tower/MC/Coldplate = 237/32.0/194 mK, Still = 1.46 K / 0.130 mbar / 0 mW, G1/G2= 74/71 mbar, 1K Pot = 1.68 K / 45±2 mbar.
1500 Tower/MC/Coldplate = 284/32.8/195 mK
1515 Tower/MC/Coldplate = 338/32.8/192 mK
Tower temp continues to rise but it looks like it's leveling off
1520 MB = 89.5%
1525 Tower peaks at 343 mK, then begins to fall
1540 Tower/MC/Coldplate = 319/33.7/186 mK
1545 Matt exits lab
1620 Tower = 182 mK and settling
1625 Matt Fritts body temp = 96.5 °F
Let's take some 0-Volt data. We can compare the traces to Run 67 data. And we can try to process it, get that set up.
1634 Start 5min series of 0V data, with 1000 beginning randoms, no trigger, 5 Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).

Next Day